Literature DB >> 15202818

First clinical trial for a new CRRT machine: the Prismaflex.

G Salvatori1, Z Ricci, M Bonello, R Ratanarat, V D'Intini, A Brendolan, M Dan, P Piccinni, R Bellomo, C Ronco.   

Abstract

A new CRRT machine has been designed to fulfill the expectations of nephrologists and intensivists operating in the common ground of critical care nephrology. The new equipment is called "Prismaflex" (Gambro-Dasco, Mirandola, Modena) and it is the natural evolution of the Prisma machine that has been utilized worldwide for CRRT in the last decade. We performed a preliminary "alfa trial" to establish usability, flexibility and realiability of the new device. Accuracy was also tested by recording various operational parameters during different intermittent and continuous renal replacement modalities. Forty-one runs were conducted on 13 patients and the difference between delivered and prescribed parameters was always lower than 2%. We concluded that the new Prismaflex is a well designed new machine for CRRT and can be safely and effectively utilized in the critical care nephrology setting.

Entities:  

Mesh:

Year:  2004        PMID: 15202818     DOI: 10.1177/039139880402700509

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  1 in total

1.  One-year safe use of the Prismaflex HF20(®) disposable set in infants in 220 renal replacement treatment sessions.

Authors:  Siegfried Rödl; Ingrid Marschitz; Christoph J Mache; Martin Koestenberger; Georg Madler; Thomas Rehak; Gerfried Zobel
Journal:  Intensive Care Med       Date:  2011-02-19       Impact factor: 17.440

  1 in total