Susan J Thorpe1, Dawn Sands, Alan B Heath, Malcolm S Hamilton, Sheena Blackmore, Trevor Barrowcliffe.
Abstract
Folate measurements, particularly for whole blood, show wide inter-laboratory and inter-methodology variability. This variability appears to be due in part to the lack of internationally accepted reference materials. A whole blood haemolysate, lyophilised in ampoules and designated 95/528, was therefore evaluated by 15 laboratories in five countries for its suitability as an International Standard (IS) for whole blood folate. The preparation was assayed using a variety of microbiological and protein-binding methodologies against local standards and calibrators. A consensus folate content was assigned to 95/528. The inclusion of three whole blood samples in the study with widely differing folate levels demonstrated a considerable reduction in inter-laboratory variability when the folate content of the samples was determined relative to the proposed IS 95/528 rather than to laboratories' local standards and calibrators. Accelerated degradation studies indicated that the folate content of 95/528 is stable when stored at -20 degrees C. On the basis of the results presented here, the World Health Organization Expert Committee on Biological Standardization established 95/528 as an IS for whole blood folate.Entities: