A no-moving-part planar array infrared spectrograph (PA-IR) equipped with a 256 x 256 mercury cadmium telluride (MCT) focal plane array has been designed and constructed. The performance of the instrument, whose frequency range extends from 2000-975 cm(-1), has been assessed in terms of resolution, bandwidth, and signal-to-noise ratio. The PA-IR spectrograph is able to record spectra with an 8.7 ms time resolution and has peak-to-peak noise levels as low as 2.4 x 10(-4) A.U. As a demonstration of the potential of PA-IR, the dynamics of reorientation of a liquid crystalline sample exposed to a single electric field pulse has been studied. It was shown that PA-IR can be used for the simultaneous acquisition of two orthogonally polarized spectra. The advantages and limitations of PA-IR, step-scan Fourier transform infrared (FT-IR), and ultrarapid-scanning FT-IR for real-time studies of reversible and irreversible phenomena are thoroughly discussed.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15198813 DOI: 10.1366/000370204873060
Source DB: PubMed Journal: Appl Spectrosc ISSN: 0003-7028 Impact factor: 2.388