S K Aggarwal1, M Kinter, D A Herold.
Abstract
An isotope dilution gas chromatography-mass spectrometry method for selenium determination in urine using 76Se as an internal standard is described. Three different derivatizing reagents, 4-nitro-o-phenylenediamine (NPD), 3,5-dibromo-o-phenylenediamine (DBPD), and 4-trifluoromethyl-o-phenylenediamine (TFMPD) were investigated for their gas chromatographic behavior including memory and precision and accuracy of isotope ratio measurements. By these criteria, the performance of these reagents was TFMPD greater than DBPD greater than NPD. Overall precision values of 1 to 7% were observed in determining Se isotope ratios at the 10-ng level, with no significant difference in using any of the three reagents. Memory effect was observed in the order NPD greater than DBPD greater than TFMPD with TFMPD showing no measurable memory effect. Accuracy of the GC-MS method was verified by quantitation of selenium in the NIST freeze-dried urine reference material SRM 2670.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1519765 DOI: 10.1016/0003-2697(92)90120-v