Moshi Geso1, Trevor Ackerly, William Patterson.
Abstract
The main advantage of radiochromic-film dosimeters is the coupling of rapid full planar-acquisition, high-spatial resolution and dose linearity. Their main limitation, however, is their low radiation sensitivity. This precludes their application to measuring doses below a few Gy. Radiochromic films are sensitive to ultraviolet radiation. In this note the results of exposing radiochromic films to x rays through an UV imaging cassette, which converts and amplifies x rays to UV radiation, are presented. These results indicate a clear increase (around 50%) in radiochromic film's sensitivity to MV x rays (6, 10, and 18 MV) when exposed through the UV phosphor.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15191286 DOI: 10.1118/1.1689012
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071