An interfering component in cardiac troponin I immunoassays-Its nature and inhibiting effect on the binding of antibodies against different epitopes.

Susann Eriksson1, Miia Junikka, Kim Pettersson.   

Abstract

OBJECTIVES: We recently reported on the occurrence of a common interfering factor (IF) that negatively affects determinations of cardiac troponin I (cTnI). The aim of the present investigation was to extend our initial finding by a detailed epitope-based determination of the location of IF and to reveal the approximate size and characteristics of IF. DESIGN AND METHODS: Two-site immunoassays using combinations of 16 monoclonal and 2 polyclonal cTnI antibodies and 1 monoclonal troponin C (TnC) antibody were used to measure the analytical recovery of cTnI or cTnI-TnC in serum samples. Gel filtration of serum samples containing IF was performed and the analytical recovery of cTnI in the fractions was determined. EDTA-plasma samples to which cTnI had been added and serum samples containing endogenous cTnI were also separated by gel filtration.
RESULTS: The mean analytical recoveries of cTnI were 28.3% (range 7.5-55.1%) and of cTnI-TnC were 23.5% (range 8.7-51.8%) in samples containing IF when antibodies against midfragment epitopes of cTnI were used. The mean recovery of cTnI was 65.1% and 73.3% for antibodies with N- and C-terminal epitopes. Gel filtration of samples with low recovery of cTnI showed that the approximate molecular mass of IF was 50-200 kDa and that the cTnI elution profiles of samples containing IF were shifted towards higher molecular mass compared with samples with less IF.
CONCLUSIONS: Antibodies against midfragment epitopes of cTnI are affected by IF to a considerable but variable extent, and the presence of IF can be demonstrated by gel filtration.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15183295     DOI: 10.1016/j.clinbiochem.2004.01.007

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  2 in total

1.  Multiple immunoassay systems are negatively interfered by circulating cardiac troponin I autoantibodies.

Authors:  Gusheng Tang; Yu Wu; Weiguo Zhao; Qian Shen
Journal:  Clin Exp Med       Date:  2011-06-08       Impact factor: 3.984

Review 2.  Evaluation of Molecularly Imprinted Polymers for Point-of-Care Testing for Cardiovascular Disease.

Authors:  Brian Regan; Fiona Boyle; Richard O'Kennedy; David Collins
Journal:  Sensors (Basel)       Date:  2019-08-09       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.