A modified internal lock-mass method for calibration of the product ions derived from sustained off-resonance irradiation collision-induced dissociation using a Fourier transform mass spectrometer.

Lifang Duan1, T-W Dominic Chan.   

Abstract

A modified internal lock-mass calibration method is introduced for improving the mass measurement accuracy of the product ion spectra derived from sustained off-resonance irradiation collision-induced dissociation (SORI-CID) Fourier transform ion cyclotron resonance (FTICR) mass spectrometry. This method involves an initial external calibration of the Fourier transform mass spectrometer to obtain the initial A- and B-terms for the equation (f(i) = A/(m/z)(i) + B). The A-term is adjusted by using an empirical relationship between the up-shift of the A-term and the pulse-gas duration, whereas the B-term is adjusted by using the mass of the unfragmented precursor ion from the SORI-CID mass spectrum of the unknown sample as internal lock-mass. These adjusted A- and B-terms are then used to provide exact mass SORI-CID calibration for the unknown sample. The modified internal lock-mass method achieved average mass measurement accuracy of approximately 3 ppm which is significantly better than that of the conventional internal lock-mass calibration ( approximately 9 ppm) and is approaching that of the internal calibration ( approximately 2 ppm) and requires no addition of internal calibrant or instrumental modifications. Copyright 2004 John Wiley & Sons, Ltd.

Mesh:

Substances:

Year:  2004        PMID: 15174182     DOI: 10.1002/rcm.1478

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  1 in total

1.  Informatics strategies for large-scale novel cross-linking analysis.

Authors:  Gordon A Anderson; Nikola Tolic; Xiaoting Tang; Chunxiang Zheng; James E Bruce
Journal:  J Proteome Res       Date:  2007-08-03       Impact factor: 4.466

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.