Literature DB >> 15162921

Determination of (+)-3,3',5-triiodo-L-thyronine (L-T3) from serum using a sequential injection analysis/immunosensor system.

Raluca-Ioana Stefan1, Jacobus Frederick van Staden, Hassan Y Aboul-Enein.   

Abstract

A sequential injection analysis/immunosensor system is proposed for the analysis of T3 in serum with a rate of 75 samples/hr. The immunosensor design is based on the physical immobilization of anti-T3 in carbon paste. The working concentration range of the immunosensor in a sequential injection analysis system is between 3.4 and 340 ng/mL with a limit of detection of 2.19 ng/mL. The system is very reliable and very easy to design and operate.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15162921     DOI: 10.1081/ias-120030527

Source DB:  PubMed          Journal:  J Immunoassay Immunochem        ISSN: 1532-1819


  1 in total

Review 1.  Developing Biosensors in Developing Countries: South Africa as a Case Study.

Authors:  Ronen Fogel; Janice Limson
Journal:  Biosensors (Basel)       Date:  2016-02-02