Literature DB >> 15150043

A biomechanical comparison of initial soft tissue tibial fixation devices: the Intrafix versus a tapered 35-mm bioabsorbable interference screw.

David N M Caborn1, Jeff C Brand, John Nyland, Yavuz Kocabey.   

Abstract

BACKGROUND: Biomechanical testing of the Intrafix device has not been performed using human tibiae. HYPOTHESIS: The Intrafix device would provide comparable or superior tibial fixation of a quadrupled hamstring tendon graft to a 35-mm-long bioabsorbable interference screw. STUDY
DESIGN: In vitro, biomechanical study.
METHODS: Eight paired human tibiae and 16 quadrupled hamstring tendon grafts were divided into 2 groups. Each quadrupled hamstring tendon graft was fixed in a tunnel sized to 0.5 mm graft diameter with either an Intrafix device or a screw.
RESULTS: Displacement at failure was greater in the Intrafix group (17.3 +/- 4.6 mm versus 10.9 +/- 4.4 mm, P =.002). Load at failure (796 +/- 193 N versus 647 +/- 269 N), stiffness (49.2 +/- 21.9 N/mm versus 64.5 +/- 22 N/mm), and bone mineral density (0.74 +/- 0.15 gm/cm(3) versus 0.74 +/- 0.14 gm/cm(3)) did not display significant differences for the Intrafix device and the screw, respectively (P >.05).
CONCLUSIONS: Displacement at failure was greater for the Intrafix device. CLINICAL RELEVANCE: Increased displacement at failure for the Intrafix group suggests slippage from sheath channel deployment. Concentric fixation may not occur when less than optimal tibial bone mineral density increases the difficulty of attaining precise sheath deployment and quadrupled hamstring tendon graft strand alignment.

Entities:  

Mesh:

Year:  2004        PMID: 15150043     DOI: 10.1177/0363546503261696

Source DB:  PubMed          Journal:  Am J Sports Med        ISSN: 0363-5465            Impact factor:   6.202


  4 in total

1.  Preparation techniques for all-inside ACL cortical button grafts: a biomechanical study.

Authors:  Raul Mayr; Christian Heinz Heinrichs; Martin Eichinger; Vinzenz Smekal; Werner Schmoelz; René Attal
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-04-19       Impact factor: 4.342

2.  Resorbable screw and sheath versus resorbable interference screw and staples for ACL reconstruction: a comparison of two tibial fixation methods.

Authors:  Christian Carulli; Fabrizio Matassi; Stefano Soderi; Luigi Sirleo; Giovanni Munz; Massimo Innocenti
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-04-27       Impact factor: 4.342

3.  Biomechanical Comparison of a Novel Tibial Fixation Technique Versus Interference Screw Fixation for ACL Reconstruction Using Soft Tissue Grafts.

Authors:  Tinghan Sun; Hangzhou Zhang
Journal:  Orthop J Sports Med       Date:  2021-08-25

4.  Can tape-screw fixation of a quadrupled semitendinosus graft in a full-length tibial tunnel provide superior fixation compared with a doubled semitendinosus-gracilis held with an interference screw? A matched-pair cadaveric biomechanical comparison.

Authors:  Christopher J Vertullo; Joseph Cadman; Dané Dabirrahmani; Richard Appleyard
Journal:  J Orthop Traumatol       Date:  2018-08-21
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.