Literature DB >> 1514718

The esophageal detector device: a rapid and accurate method for assessing tracheal versus esophageal intubation in a porcine model.

R G Foutch1, M D Magelssen, J G MacMillan.   

Abstract

STUDY
OBJECTIVES: To assess time and accuracy of the esophageal detector device (EDD), disposable end-tidal CO2 monitor (ETCO2), and standard clinical methods for detection of endotracheal tube placement.
DESIGN: Prospective, randomized, single-blinded, controlled laboratory investigation.
METHODS: Thirty airway managers (physicians, nurse anesthetists, and paramedics) used one pig (Sus scrofa) as the intubated, respiratory depressed/arrest model.
INTERVENTIONS: Part 1: A standard 7.5-mm endotracheal tube was placed in either the esophagus or the trachea of the anesthetized swine. Anatomic location was verified by bronchoscopy. Airway managers blinded to the endotracheal tube location were assigned randomly to identify tube position by one of three methods (EDD, ETCO2, or clinical methods). Speed and accuracy of the assessment were recorded. Part 2: A second identical tube was placed, so that both the esophagus and the trachea were intubated; then, the esophageal tube was bag-ventilated for one minute. Each blinded airway manager, using only the EDD, determined placement site of both tubes.
RESULTS: Part 1: Mean time to determine tube placement for group A (EDD) was 13.8 seconds; group B (ETCO2), 31.5 seconds; and group C (clinical methods), 39 seconds. Comparison by analysis of variance yielded a value of P less than .001. Both groups A and B were 100% accurate, whereas 30% of the subjects from group C mistakenly assessed an esophageal tube as in the trachea. Part 2: The EDD remained 100% sensitive and specific despite prior ventilation of the esophageal tube.
CONCLUSION: In this porcine model, the EDD and ETCO2 were more accurate than clinical methods in determining endotracheal tube placement. The EDD demonstrated a significant time advantage over both ETCO2 and clinical methods. Prior ventilation of the esophageal tube does not interfere with the accuracy of the EDD.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1514718     DOI: 10.1016/s0196-0644(05)80647-x

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  3 in total

1.  Predictive Value of Tracheal Rapid Ultrasound Exam Performed in the Emergency Department for Verification of Tracheal Intubation.

Authors:  Babak Masoumi; Reza Azizkhani; Gilava Hedayati Emam; Morteza Asgarzadeh; Behrouz Zargar Kharazi
Journal:  Open Access Maced J Med Sci       Date:  2017-06-17

2.  Anesthesiology and the difficult airway - Where do we currently stand?

Authors:  Martina Richtsfeld; Kumar G Belani
Journal:  Ann Card Anaesth       Date:  2017 Jan-Mar

3.  Screening Characteristics of Bedside Ultrasonography in Confirming Endotracheal Tube Placement; a Diagnostic Accuracy Study.

Authors:  Hamid Zamani Moghadam; Mohamad Davood Sharifi; Hasan Rajabi; Mojtaba Mousavi Bazaz; Ali Alamdaran; Niazmohammad Jafari; Seyed Amir Masoud Hashemian; Morteza Talebi Deloei
Journal:  Emerg (Tehran)       Date:  2017-01-09
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.