Literature DB >> 15140727

Anatomically conformable, three-dimensional, detachable platinum microcoil system for the treatment of intracranial aneurysms.

David Niemann1, Richard Aviv, Caroline Cowsill, Mary Sneade, Andrew J Molyneux.   

Abstract

BACKGROUND AND
PURPOSE: The configurations of detachable platinum coils have evolved over the last decade. Our objective was to assess the efficacy of the anatomically conformable three-dimensional (ACT) detachable platinum microcoil in the treatment of intracranial aneurysms.
METHODS: A series of 141 patients presented between May 2000 and August 2002 to a single neurointerventionalist for endovascular treatment of intracranial aneurysms. Eight patients were treated at another institute or are now overseas. The remaining 133 patients had 141 aneurysms, which were treated; 110 were ruptured and 31 were unruptured. The clinical conditions and aneurysm locations were recorded. The procedural technique was the same as that used with the Guglielmi detachable coil (GDC) device, though the coil detachment mechanism was unique. The ACT microcoil was used in almost all cases as the initial coil.
RESULTS: Complete (100%) occlusion was achieved in 104 (74%) aneurysms. Thirty four (24%) had subtotal occlusion > or = 95%), and three (2%) had incomplete occlusion (<95%). The ACT coil conformed to both spherical and irregularly shaped aneurysms. No procedure-related deaths occurred, and there was no aneurysm rupture related to coil deployment. No postprocedural rebleeding was observed.
CONCLUSION: The ACT microcoil device is feasible as an alternative platinum coil device, offering some advantages over other currently available devices.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15140727      PMCID: PMC7974493     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  13 in total

1.  Unruptured intracranial aneurysms--risk of rupture and risks of surgical intervention.

Authors: 
Journal:  N Engl J Med       Date:  1998-12-10       Impact factor: 91.245

2.  International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial.

Authors:  Andrew Molyneux; Richard Kerr; Irene Stratton; Peter Sandercock; Mike Clarke; Julia Shrimpton; Rury Holman
Journal:  Lancet       Date:  2002-10-26       Impact factor: 79.321

3.  Endovascular treatment of giant and large intracranial aneurysms by using a combination of stent placement and liquid polymer injection.

Authors:  Michel E Mawad; Saruhan Cekirge; Elisa Ciceri; Isil Saatci
Journal:  J Neurosurg       Date:  2002-03       Impact factor: 5.115

4.  Report of World Federation of Neurological Surgeons Committee on a Universal Subarachnoid Hemorrhage Grading Scale.

Authors: 
Journal:  J Neurosurg       Date:  1988-06       Impact factor: 5.115

5.  Treatment of an intracranial aneurysm using a new three-dimensional-shape Guglielmi detachable coil: technical case report.

Authors:  A M Malek; R T Higashida; C C Phatouros; C F Dowd; V V Halbach
Journal:  Neurosurgery       Date:  1999-05       Impact factor: 4.654

6.  Electrothrombosis of saccular aneurysms via endovascular approach. Part 2: Preliminary clinical experience.

Authors:  G Guglielmi; F Viñuela; J Dion; G Duckwiler
Journal:  J Neurosurg       Date:  1991-07       Impact factor: 5.115

7.  Electrothrombosis of saccular aneurysms via endovascular approach. Part 1: Electrochemical basis, technique, and experimental results.

Authors:  G Guglielmi; F Viñuela; I Sepetka; V Macellari
Journal:  J Neurosurg       Date:  1991-07       Impact factor: 5.115

8.  Intracranial berry aneurysms: angiographic and clinical results after endovascular treatment.

Authors:  C Cognard; A Weill; L Castaings; A Rey; J Moret
Journal:  Radiology       Date:  1998-02       Impact factor: 11.105

9.  Embolisation of recently ruptured intracranial aneurysms.

Authors:  J V Byrne; A J Molyneux; R P Brennan; S A Renowden
Journal:  J Neurol Neurosurg Psychiatry       Date:  1995-12       Impact factor: 10.154

10.  Endovascular occlusion of intracranial aneurysms with electrically detachable coils: correlation of aneurysm neck size and treatment results.

Authors:  A Fernandez Zubillaga; G Guglielmi; F Viñuela; G R Duckwiler
Journal:  AJNR Am J Neuroradiol       Date:  1994-05       Impact factor: 3.825

View more
  6 in total

1.  Cerecyte coils in the treatment of intracranial aneurysms: a preliminary clinical study.

Authors:  M Bendszus; L Solymosi
Journal:  AJNR Am J Neuroradiol       Date:  2006 Nov-Dec       Impact factor: 3.825

2.  Safety and clinical efficacy of Onyx for embolization of extracranial head and neck vascular anomalies.

Authors:  R Thiex; I Wu; J B Mulliken; A K Greene; R Rahbar; D B Orbach
Journal:  AJNR Am J Neuroradiol       Date:  2011-03-31       Impact factor: 3.825

3.  Embolization of Cerebral Aneurysms with Spherically Shaped Detachable Microcoils (MicruSphere Microcoil System). A Single Centre Experience.

Authors:  D H Lee; A Arat; H Morsi; L-D Jou; M E Mawad
Journal:  Interv Neuroradiol       Date:  2009-04-15       Impact factor: 1.610

Review 4.  Delayed coil migration from a small wide-necked aneurysm after stent-assisted embolization: case report and literature review.

Authors:  Bu-Lang Gao; Ming-Hua Li; Yong-Li Wang; Chun Fang
Journal:  Neuroradiology       Date:  2006-04-06       Impact factor: 2.804

5.  Endovascular treatment of intracranial aneurysms with bioactive Cerecyte coils: effects on treatment stability.

Authors:  Serdar Geyik; Kivilcim Yavuz; Onur Ergun; Osman Koc; Saruhan Cekirge; Isil Saatci
Journal:  Neuroradiology       Date:  2008-05-16       Impact factor: 2.804

6.  Sapphire platinum detachable coil experience in a tertiary-care facility.

Authors:  Abutaher M Yahia; Vickie Gordon; John Whapham; Ali Malek; Mohammed Rehman; Richard D Fessler
Journal:  Neurocrit Care       Date:  2007       Impact factor: 3.210

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.