Improved lung edema monitoring with coronary vein pacing leads: a simulation study.

Andres Belalcazar1, Robert P Patterson.   

Abstract

This computer simulation study compared the ability of left ventricular coronary vein (LV) pacemaker leads against right ventricular (RV) and right atrial (RA) leads to monitor lung edema using electrical impedance measurements. MRI images were used to construct electrical models of the thorax. Four lead configurations were tested with increases of pulmonary edema, intravascular fluids and heart dilation. The impedance changes observed at end systole with severe lung edema were 8.5%, 11.2%, 12.3% and 26.8% for the RA, RV, RV coil and LV configurations, respectively. Sensitivities in ohms per litre of lung fluid were 19.15, 19.15, 25.07 and 52.11 for the same configurations. The impedance changes for intravascular fluid overload with constant lung status were 1%, 1.3%, 9.2% and 6.4% while the sensitivities were 2, 2, 17 and 11 ohms per litre of intravascular fluid, respectively. Regional analysis of the thoracic sources of impedance revealed a high sensitivity near pacing electrodes and generator, and a low sensitivity to the right lung and all pulmonary vessels. Simulations showed that LV leads have a threefold advantage in sensitivity when monitoring lung edema in comparison to conventional RV leads. To monitor vascular and lung fluids independently, combined impedance configurations may be used. Regional sensitivities must be taken into account for proper clinical interpretation of impedance changes.

Entities:  

Mesh:

Year:  2004        PMID: 15132313     DOI: 10.1088/0967-3334/25/2/007

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  5 in total

Review 1.  How to improve outcomes: should we put more emphasis on programming and medical care and less on patient selection?

Authors:  Laszlo Buga
Journal:  Heart Fail Rev       Date:  2012-11       Impact factor: 4.214

2.  Device monitoring strategies in acute heart failure syndromes.

Authors:  Michael A Samara; W H Wilson Tang
Journal:  Heart Fail Rev       Date:  2011-09       Impact factor: 4.214

3.  Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors.

Authors:  Dirar S Khoury; Mihir Naware; Jeff Siou; Andreas Blomqvist; Nilesh S Mathuria; Jianwen Wang; Hue-Teh Shih; Sherif F Nagueh; Dorin Panescu
Journal:  J Am Coll Cardiol       Date:  2009-03-24       Impact factor: 24.094

4.  Safety and efficacy aspects of pulsed field ablation catheters as a function of electrode proximity to blood and energy delivery method.

Authors:  Andres Belalcazar
Journal:  Heart Rhythm O2       Date:  2021-10-13

5.  Analysis of Consistency of Transthoracic Bioimpedance Measurements Acquired with Dry Carbon Black PDMS Electrodes, Adhesive Electrodes, and Wet Textile Electrodes.

Authors:  Hugo F Posada-Quintero; Natasa Reljin; Caitlin Eaton-Robb; Yeonsik Noh; Jarno Riistama; Ki H Chon
Journal:  Sensors (Basel)       Date:  2018-05-26       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.