| Literature DB >> 15125142 |
R Modre1, B Tilg, G Fischer, F Hanser, B Messnarz, M Seger, F Hintringer, F X Roithinger.
Abstract
Non-invasive imaging of cardiac electrophysiology provides a non-invasive way of obtaining information about electrical excitation. An iterative algorithm based on a general regularisation scheme for non-linear, ill-posed problems in Hilbert scales was applied to the electrocardiographic inverse problem, imaging the ventricular surface activation time (AT) map. This method was applied to electrocardiographic data from a 31-year-old healthy volunteer and a 24-year-old patient suffering from a Wolff-Parkinson-White (WPW) syndrome. The objective was to evaluate non-invasive AT imaging of an autonomous sinus rhythm and to quantify the localisation error of non-invasive AT imaging by localising the accessory pathway of the WPW syndrome and a pacing site for left ventricle pacing. The distances between the invasive and non-invasive localisation of the pacing site and the accessory pathway were 8 mm and 5 mm. The clinical case presented, shows that this non-invasive AT imaging approach may enable the reconstruction of single focal events with sufficient accuracy for potential clinical application.Entities:
Mesh:
Year: 2004 PMID: 15125142 DOI: 10.1007/bf02344624