Value and limitations of aortic valve resistance with particular consideration of low flow-low gradient aortic stenosis: an in vitro study.

Julia Mascherbauer1, Heinrich Schima, Raphael Rosenhek, Martin Czerny, Gerald Maurer, Helmut Baumgartner.   

Abstract

BACKGROUND: The calculation of valve resistance (R) rather than aortic valve area (AVA) has been proposed for the assessment of aortic stenosis (AS), based on the claim that it is less flow-dependent. Even more importantly, valve resistance has been reported to distinguish between truly severe and "pseudosevere" AS in patients with low cardiac output. However, the diagnostic value of valve resistance remains controversial. METHODS AND
RESULTS: Models of stenotic aortic valves (plates and nozzles) and biological stenotic valves were studied in a pulsatile in vitro circuit using Doppler ultrasound and direct pressure and flow measurements. Anatomic AVAs ranged from 0.5 to 1.25 cm2; cardiac output varied from 1.8 to 9.0 l/min. Effective AVA was calculated with the continuity equation. The orifices of the biological valves were recorded with a video camera for planimetry. In low flow-low gradient AS, truly severe stenosis was defined by an AVA remaining <0.85 cm2 after flow normalisation, whereas AVA increased beyond 0.85 cm2 in pseudosevere AS. In rigid stenoses, valve resistance increased significantly with flow, while in bioprostheses this flow dependence was partially masked by an actual increase of the anatomic orifice area. In low flow-low gradient AS, valve resistance was significantly smaller in pseudosevere AS compared to truly severe AS (129 +/- 28 vs. 176 +/- 33 dyne s cm(-5); p<0.001) at a similar baseline effective AVA. After the exclusion of datasets with mean gradients <15 and >35 mmHg, the difference in valve resistance between truly severe and pseudosevere AS was no longer significant (162 +/- 26 vs. 141 +/- 22 dyne s cm(-5); p=0.08). Nevertheless, valve resistance <120 dyne s cm(-5) was found only in pseudosevere stenoses while valve resistance >180 dyne s cm(-5) marked truly severe stenosis.
CONCLUSIONS: Valve resistance is flow-dependent and not superior to calculated AVA for the assessment of AS. In low flow-low gradient AS, valve resistance <120 dyne s cm(-5) identifies pseudosevere AS, whereas valve resistance >180 dyne s cm(-5) implies truly severe AS. However, values between 120 and 180 dyne s cm(-5) are nondiagnostic, requiring repeated AVA calculations after flow normalisation.

Entities:  

Mesh:

Year:  2004        PMID: 15120890     DOI: 10.1016/j.ehj.2004.02.011

Source DB:  PubMed          Journal:  Eur Heart J        ISSN: 0195-668X            Impact factor:   29.983


  4 in total

1.  Assessment of aortic stenosis severity: check the valve but don't forget the arteries!

Authors:  Philippe Pibarot; Jean G Dumesnil
Journal:  Heart       Date:  2007-07       Impact factor: 5.994

2.  Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves.

Authors:  Ovandir Bazan; Jayme Pinto Ortiz
Journal:  Braz J Cardiovasc Surg       Date:  2016-04

3.  Ejection fraction-velocity ratio for the assessment of aortic bioprosthetic valves in patients with systolic dysfunction.

Authors:  P Cattaneo; P Marchetti; M Baravelli; A Rossi; G Mariscalco; S Ghiringhelli; C Anzà
Journal:  Can J Cardiol       Date:  2009-03       Impact factor: 5.223

4.  Evaluation of the intraoperative specimens of the thoracic and abdominal aorta.

Authors:  Andrzej Juraszek; Günther Bayer; Tomasz Dziodzio; Artur Kral; Günther Laufer; Marek Ehrlich
Journal:  J Cardiothorac Surg       Date:  2013-04-24       Impact factor: 1.637

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.