Comparison of cast Ti-Ni alloy clasp retention with conventional removable partial denture clasps.

DongSuk Kim1, ChanJin Park, YangJin Yi, LeeRa Cho.   

Abstract

STATEMENT OF PROBLEM: The elastic property of Ti-Ni alloy may be a desirable property for cast removable partial dentures. However, little is known about the retentive properties of cast Ti-Ni alloy clasps.
PURPOSE: This in vitro study investigated the retentive force of various types of clasps during repeated cycles of placement and removal to determine whether Ti-Ni alloy clasps maintain their initial retentive force under varied conditions, including different retentive undercut depths and clasp size (thickness and width).
MATERIAL AND METHODS: The test models were developed using a mandibular dentiform with a horizontal plane of occlusion. Two complete metal abutment crowns were made for the left second molar, differing only in retentive undercut depths. A total of 98 clasps (n=7) were fabricated, including 14 wrought wire clasps and 28 clasps from each of the following alloys: Co-Cr alloy (Biosil), Type IV gold alloy (Cast-4), and Ti-Ni alloy (TN-10). Clasps were made to engage one of 2 retentive undercut depths (0.25 mm or 0.75 mm) and were made in 2 sizes (0.8 mm or 1.4 mm). Each clasp was radiographically examined for casting defects and porosity. The force (N) required to remove the clasps was measured using a universal testing machine with a crosshead speed of 10 mm/min. After measuring the retentive force, a masticatory simulator was used to cycle the clasps on and off the metal crown 500 times, simulating the insertion and removal of a removable partial denture clasp. The cycling sequence was repeated 10 times, and retention force was measured after each sequence. The internal surface of the clasp arms was examined for evidence of metal fatigue using scanning electron microscopy. Data were subjected to 2-way and 4-way analysis of variance, followed by Scheffe's multiple comparison test (alpha=.05).
RESULTS: Co-Cr alloy and gold alloy clasps in the 0.25-mm retentive undercut groups experienced a gradual decrease in retentive force measurements (P<.001). In contrast, the Ti-Ni alloy clasps maintained a retentive force of approximately 1.8 N and 2.6 N for the 0.8 mm and 1.4 mm clasp groups, respectively. Although Ti-Ni alloy clasps' retentive force was reduced slightly after the first cycling sequence, it was negligible compared with other clasps tested. The wrought wire clasps also retained their retentive force until the final cycling sequence. A similar trend was found in the clasp groups engaging 0.75-mm retentive undercuts.
CONCLUSION: Although the end-point retention for all the clasps was similar, there was less change in the retentive force of the cast Ti-Ni alloy clasps after repeated cycling sequences of simulated placement and removal.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15116040     DOI: 10.1016/j.prosdent.2004.02.015

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  6 in total

1.  Comparison of joint designs for laser welding of cast metal plates and wrought wires.

Authors:  Yasuko Takayama; Rie Nomoto; Hiroyuki Nakajima; Chikahiro Ohkubo
Journal:  Odontology       Date:  2011-11-12       Impact factor: 2.634

2.  Comparative study of two commercially pure titanium casting methods.

Authors:  Renata Cristina Silveira Rodrigues; Adriana Claudia Lapria Faria; Iara Augusta Orsi; Maria da Gloria Chiarello de Mattos; Ana Paula Macedo; Ricardo Faria Ribeiro
Journal:  J Appl Oral Sci       Date:  2010 Sep-Oct       Impact factor: 2.698

3.  The effect of environment (Dry and Natural Saliva) on clasp retention: In vitro study - Part I.

Authors:  Bilal Mourshed; Nashwan Mohammed Qaed; Hashem Motahir Al-Shamiri; Nader Alaizari; Saleh Sulaiman Alhamdah; Amal Alfaqih
Journal:  Eur J Dent       Date:  2017 Jul-Sep

4.  Retention Forces of Prosthetic Clasps over a Simulated Wearing Period of Six Years In-Vitro: Direct Metal Laser Melting Versus Dental Casting.

Authors:  Moritz Mutschler; Florian Schweitzer; Sebastian Spintzyk; Jürgen Geis-Gerstorfer; Fabian Huettig
Journal:  Materials (Basel)       Date:  2020-11-25       Impact factor: 3.623

5.  Effect of clasp type and pullout location on clasp retention in different environment: In vitro study.

Authors:  Bilal Mourshed; Fuad Abdo Al-Sabri; Nashwan Ahmed Qaed; Nader Alaizari; Hashem Motahir Al-Shamiri; Amal Alfaqih
Journal:  Eur J Dent       Date:  2017 Apr-Jun

6.  Examination of inter-rater and intra-rater reliability during retentive force measurement of different clasps using the developed small-sized retentive force measurement device.

Authors:  Hitoshi Akiyama; Maiko Sakamoto; Ryoichi Akama; Jun Takei
Journal:  BMC Oral Health       Date:  2020-08-20       Impact factor: 2.757

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.