Influence of occlusal forces on stress distribution in preloaded dental implant screws.

Ibrahim Alkan1, Atilla Sertgöz, Bülent Ekici.   

Abstract

STATEMENT OF PROBLEM: Abutment and prosthetic loosening of single and multiple screw-retained, implant-supported fixed partial dentures is a concern.
PURPOSE: The purpose of this study was to investigate stress distribution of preloaded dental implant screws in 3 implant-to-abutment joint systems under simulated occlusal forces.
MATERIAL AND METHODS: Three abutment-to-implant joint systems were simulated by using the 3-dimensional finite element analysis method: (1) Branemark external hexagonal screw-retained abutment, (2) ITI 8-degree Morse tapered cemented abutment, and (3) ITI 8-degree Morse tapered plus internal octagonal screw-retained abutment. A thermal load and contact analysis method were used to simulate the preload resulting from the manufacturers' recommended torques in implant screw joint assemblies. The simulated preloaded implants were then loaded with 3 simulated static occlusal loads (10 N; horizontal, 35 N; vertical, 70 N; oblique) on the crown position onto the implant complex.
RESULTS: Numeric and graphical results demonstrated that the stresses increased in both the abutment and prosthetic screws in the finite element models after simulated horizontal loading. However, when vertical and oblique static loads were applied, stresses decreased in the external hexagonal and internal octagonal plus 8-degree Morse tapered abutment and prosthetic screws with the exception of the prosthetic screw of ITI abutment after 70-N oblique loading. Stresses increased in the ITI 8-degree Morse tapered cemented abutment after both vertical and oblique loads.
CONCLUSION: Although an increase or decrease was demonstrated for the maximum calculated stress values in preloaded screws after occlusal loads, these maximum stress values were well below the yield stress of both abutment and prosthetic screws of 2 implant systems tested. The results imply that the 3 implant-to-abutment joint systems tested may not fail under the simulated occlusal forces.

Mesh:

Substances:

Year:  2004        PMID: 15116032     DOI: 10.1016/j.prosdent.2004.01.016

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  16 in total

1.  Biomechanical analysis of the wrist arthroplasty in rheumatoid arthritis: a finite element analysis.

Authors:  M N Bajuri; Mohammed Rafiq Abdul Kadir; Malliga Raman Murali; T Kamarul
Journal:  Med Biol Eng Comput       Date:  2012-11-03       Impact factor: 2.602

2.  Nonlinear finite element analysis of three implant- abutment interface designs.

Authors:  Chun-Bo Tang; Si-Yu Liul; Guo-Xing Zhou; Jin-Hua Yu; Guang-Dong Zhang; Yi-Dong Bao; Qiu-Ju Wang
Journal:  Int J Oral Sci       Date:  2012-06       Impact factor: 6.344

3.  Survival Probability, Weibull Characteristics, Stress Distribution, and Fractographic Analysis of Polymer-Infiltrated Ceramic Network Restorations Cemented on a Chairside Titanium Base: An In Vitro and In Silico Study.

Authors:  João P M Tribst; Amanda M O Dal Piva; Alexandre L S Borges; Lilian C Anami; Cornelis J Kleverlaan; Marco A Bottino
Journal:  Materials (Basel)       Date:  2020-04-16       Impact factor: 3.623

4.  All-ceramic and porcelain-fused-to-metal fixed partial dentures: a comparative study by 2D finite element analyses.

Authors:  Andréa Barreira Motta; Luiz Carlos Pereira; Andréia R C C da Cunha
Journal:  J Appl Oral Sci       Date:  2007-10       Impact factor: 2.698

5.  Comparative evaluation of osseointegrated dental implants based on platform-switching concept: influence of diameter, length, thread shape, and in-bone positioning depth on stress-based performance.

Authors:  Luigi Baggi; Michele Di Girolamo; Giuseppe Vairo; Gianpaolo Sannino
Journal:  Comput Math Methods Med       Date:  2013-06-19       Impact factor: 2.238

6.  A three-dimensional finite element study on the stress distribution pattern of two prosthetic abutments for external hexagon implants.

Authors:  Wagner Moreira; Caio Hermann; Jucélio Tomás Pereira; Jean Anacleto Balbinoti; Rodrigo Tiossi
Journal:  Eur J Dent       Date:  2013-10

7.  Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study.

Authors:  Krishna Chaitanya Kanneganti; Dileep Nag Vinnakota; Srinivas Rao Pottem; Mahesh Pulagam
Journal:  J Indian Prosthodont Soc       Date:  2018 Apr-Jun

8.  The Influence of Custom-Milled Framework Design for an Implant-Supported Full-Arch Fixed Dental Prosthesis: 3D-FEA Sudy.

Authors:  João Paulo Mendes Tribst; Amanda Maria de Oliveira Dal Piva; Roberto Lo Giudice; Alexandre Luiz Souto Borges; Marco Antonio Bottino; Ettore Epifania; Pietro Ausiello
Journal:  Int J Environ Res Public Health       Date:  2020-06-05       Impact factor: 3.390

9.  Improvement of osteogenesis by a uniform PCL coating on a magnesium screw for biodegradable applications.

Authors:  Yu-Kyoung Kim; Kwang-Bok Lee; Seo-Young Kim; Yong-Seok Jang; Jin Hyeok Kim; Min-Ho Lee
Journal:  Sci Rep       Date:  2018-09-05       Impact factor: 4.379

10.  Mechanical Assessment of Fatigue Characteristics between Single- and Multi-Directional Cyclic Loading Modes on a Dental Implant System.

Authors:  Won Hyeon Kim; Eun Sung Song; Kyung Won Ju; Dohyung Lim; Dong-Wook Han; Tae-Gon Jung; Yong-Hoon Jeong; Jong-Ho Lee; Bongju Kim
Journal:  Materials (Basel)       Date:  2020-03-27       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.