Literature DB >> 15100946

Comparison of EUS-guided 19-gauge Trucut needle biopsy with EUS-guided fine-needle aspiration.

S Varadarajulu1, M Fraig, N Schmulewitz, S Roberts, S Wildi, R H Hawes, B J Hoffman, M B Wallace.   

Abstract

BACKGROUND AND STUDY AIMS: The accuracy of endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) depends on immediate specimen review by a cytopathologist. Stromal tumors, lymphoma, and well-differentiated pancreatic cancer are difficult to diagnose on the basis of cytology alone. To overcome these limitations, a 19-gauge Trucut needle has been developed to obtain histological samples at EUS. This pilot study compares the specimen adequacy and diagnostic accuracy of EUS-guided Trucut needle biopsy (EUS-TNB) with EUS-FNA. PATIENTS AND METHODS: A total of 18 patients underwent EUS-TNB and EUS-FNA. The specimen adequacy and diagnostic accuracy of the two techniques was compared. The technical performance and safety profile of the Trucut needle were also evaluated.
RESULTS: The EUS-TNB specimen was adequate for evaluation in 15/18 patients compared with 18/18 with EUS-FNA (83 % vs. 100 %, not significant). The diagnostic accuracy of EUS-TNB was not significantly different from EUS-FNA (78 % vs. 89 %). Two complications were encountered: one patient developed mediastinitis and required surgery; another had immediate bleeding that was managed conservatively. One technical problem was encountered: the Trucut needle failed to deploy after two passes when a gastric stromal cell tumor was being biopsied.
CONCLUSION: The diagnostic accuracy of the new EUS-TNB is comparable to that of EUS-FNA. In our experience, the overall efficacy and safety profile of the Trucut needle appears modest.

Entities:  

Mesh:

Year:  2004        PMID: 15100946     DOI: 10.1055/s-2004-814316

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  64 in total

1.  Comparison of diagnostic accuracy of endoscopic ultrasound-guided fine-needle aspiration with 22- and 25-gauge needles in the same patients.

Authors:  Mitsuhiro Kida; Masao Araki; Shiro Miyazawa; Hiroko Ikeda; Miyoko Takezawa; Hidehiko Kikuchi; Maya Watanabe; Hiroshi Imaizumi; Wasaburo Koizumi
Journal:  J Interv Gastroenterol       Date:  2011-07-01

Review 2.  Endoscopic management of gastrointestinal stromal tumors.

Authors:  Vinay Chandrasekhara; Gregory G Ginsberg
Journal:  Curr Gastroenterol Rep       Date:  2011-12

3.  Diagnosis of pancreatic neuroendocrine tumors and the role of endoscopic ultrasound.

Authors:  Linda S Lee
Journal:  Gastroenterol Hepatol (N Y)       Date:  2010-08

Review 4.  Endoscopic ultrasonography: Transition towards the future of gastro-intestinal diseases.

Authors:  Stefania De Lisi; Marc Giovannini
Journal:  World J Gastroenterol       Date:  2016-02-07       Impact factor: 5.742

5.  Comparison of 22G reverse-beveled versus standard needle for endoscopic ultrasound-guided sampling of solid pancreatic lesions.

Authors:  Abdullah Alatawi; Frédéric Beuvon; Sophie Grabar; Sarah Leblanc; Stanislas Chaussade; Benoit Terris; Maximilien Barret; Frédéric Prat
Journal:  United European Gastroenterol J       Date:  2015-08       Impact factor: 4.623

6.  Impact of endoscopic ultrasound-guided fine needle biopsy for diagnosis of pancreatic masses.

Authors:  Julio Iglesias-Garcia; Enrique Dominguez-Munoz; Antonio Lozano-Leon; Ihab Abdulkader; Jose Larino-Noia; Jose Antunez; Jeronimo Forteza
Journal:  World J Gastroenterol       Date:  2007-01-14       Impact factor: 5.742

Review 7.  Endoscopic ultrasound advances, part 1: diagnosis.

Authors:  Edward Kim; Jennifer J Telford
Journal:  Can J Gastroenterol       Date:  2009-09       Impact factor: 3.522

8.  Yields and Utility of Endoscopic Ultrasonography-Guided 19-Gauge Trucut Biopsy versus 22-Gauge Fine Needle Aspiration for Diagnosing Gastric Subepithelial Tumors.

Authors:  Hee Kyong Na; Jeong Hoon Lee; Young Soo Park; Ji Yong Ahn; Kwi-Sook Choi; Do Hoon Kim; Kee Don Choi; Ho June Song; Gin Hyug Lee; Hwoon-Yong Jung; Jin-Ho Kim
Journal:  Clin Endosc       Date:  2015-03-27

9.  Endoscopic ultrasound-guided fine-needle aspiration and Trucut biopsy in thoracic lesions: when tissue is the issue.

Authors:  Ian Storch; Mubashir Shah; R Thurer; Elio Donna; A Ribeiro
Journal:  Surg Endosc       Date:  2007-05-04       Impact factor: 4.584

10.  Ultrasound-guided vs endoscopic ultrasound-guided fine-needle aspiration for pancreatic cancer diagnosis.

Authors:  Masato Matsuyama; Hiroshi Ishii; Kensuke Kuraoka; Seigo Yukisawa; Akiyoshi Kasuga; Masato Ozaka; Sho Suzuki; Kouichi Takano; Yuko Sugiyama; Takao Itoi
Journal:  World J Gastroenterol       Date:  2013-04-21       Impact factor: 5.742

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.