Wrist blood pressure overestimates blood pressure measured at the upper arm.

Paolo Palatini1, Daniele Longo, Gianluca Toffanin, Olivo Bertolo, Vania Zaetta, Achille C Pessina.   

Abstract

OBJECTIVE: Whether blood pressure (BP) measured at the wrist differs from blood pressure measured at the arm is not well known. The aim of this study was to compare the BP readings obtained at the arm with those obtained at the forearm and to assess whether the wrist-arm discrepancies were related to subjects' clinical characteristics.
METHODS: We measured blood pressure at the forearm and at the upper arm in 85 subjects using conventional sphygmomanometry. Wrist-arm blood pressure discrepancies were assessed in relation to gender, age, body mass index, skin-fold thickness, arm size, blood pressure level, and arterial compliance measured with the HDI/Pulsewave CR-2000.
RESULTS: Blood pressure measured at the wrist consistently overestimated blood pressure taken at the arm with a mean (+/-SD) discrepancy of 8.2 +/- 9.7/9.2 +/- 6.4 mmHg. The systolic blood pressure differences were greater in men than in women (p=0.006) and, among the men, varied according to arm adiposity (positive association, p=0.01). In men, diastolic blood pressure differences correlated with diastolic blood pressure level (negative association, p=0.01). Among the women, only age (p=0.04) was a significant positive independent predictor of the wrist-arm diastolic BP differences.
CONCLUSIONS: These results indicate that forearm blood pressure measurement markedly overestimates upper arm blood pressure and that the between-site difference may vary from subject to subject. Wrist blood pressure measurement is not a valid alternative to traditional measurement at the arm and its use should be discouraged.

Entities:  

Mesh:

Year:  2004        PMID: 15096904     DOI: 10.1097/00126097-200404000-00004

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  4 in total

1.  The impact of arm position and pulse pressure on the validation of a wrist-cuff blood pressure measurement device in a high risk population.

Authors:  Ali Reza Khoshdel; Shane Carney; Alastair Gillies
Journal:  Int J Gen Med       Date:  2010-04-08

2.  Blood pressure measurement in obese patients with cone-shaped arms.

Authors:  Josiane Lima de Gusmão; Decio Mion
Journal:  J Clin Hypertens (Greenwich)       Date:  2007-02       Impact factor: 3.738

Review 3.  Blood pressure monitoring techniques in the natural state of multi-scenes: A review.

Authors:  Ziyi Liu; Congcong Zhou; Hongwei Wang; Yong He
Journal:  Front Med (Lausanne)       Date:  2022-08-26

4.  Use of a human wrist blood pressure monitor for arterial blood pressure measurements in normotensive conscious dogs in comparison to veterinary high-definition oscillometry.

Authors:  Elisa Martinelli; Riccardo Ferriani; Annamaria Zanaboni; Roberto Toschi Corneliani; Chiara Locatelli
Journal:  Vet Med Sci       Date:  2022-05-13
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.