| Literature DB >> 15095357 |
Wes E Steiner1, Brian H Clowers, William A English, Herbert H Hill.
Abstract
The use of an atmospheric pressure matrix-assisted laser desorption/ionization (AP-MALDI) source was employed with an atmospheric pressure ion mobility spectrometer (APIMS) and an orthogonal acceleration reflector time-of-flight mass spectrometer (TOFMS) to analyze dipeptide and biogenic amine mixtures from a liquid glycerol 2,5-dihydroxybenzoic acid (DHB) matrix. Improved sensitivities were obtained by the addition of a localized electrical (corona) discharge in conjunction with the AP-MALDI source. Enhanced sample ionization efficiency created by this combination provided an overall elevation in signal intensity of approximately 1.3 orders in magnitude. Combinations of three dipeptides (Gly-Lys, Ala-Lys, and Val-Lys) and nine biogenic amines (dopamine, serotonin, B-phenylethylamine, tyramine, octopamine, histamine, tryptamine, spermidine, and spermine) were resolved in less than 18 ms. In many cases, reduced mobility constants (K(o)) were determined for these analytes for the first time. Ion mobility drift times, flight times, arbitrary signal intensities, and collision-induced dissociation (CID) fragmentation product signatures are reported for each of the samples. Copyright 2004 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15095357 DOI: 10.1002/rcm.1419
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419