RATIONALE AND OBJECTIVES: The objective of this study was to investigate the potential for artifact-free coronary magnetic resonance angiography (cMRA) in the presence of dedicated metallic MR stents in vitro and in a swine model. METHODS: All investigations were performed at 1.5 T, applying a standard cMRA gradient echo sequence with a T2 preparation pulse. Two prototypes of each hand-woven, mechanically woven, and lasered Aachen Resonance Coronary MR Stents made out of an MR-compatible metallic alloy and dilated to 2.5 mm and 4 mm were examined in a water bath. RESULTS: Artifact behavior was judged independently by 2 radiologists as showing "no artifacts" for all tested stent types. Signal-to-noise ratios inside and outside of the stents were measured yielding a Pearson correlation coefficient of 0.98 (y = 1.22 + 0.92x). Nineteen stents (8 hand woven, 3 mechanically woven, 8 lasered) were deployed in coronary arteries of 19 domestic pigs and were examined by cMRA. Artifact behavior of the stents was analyzed by measuring the signal-to-noise ratio at the stent positions and compared with signal-to-noise ratio measurements outside of the stents, yielding a Pearson correlation coefficient of 0.90 (y = -0.75 + 1.06x). CONCLUSIONS: All 3 prototypes of coronary MR stents allowed complete visualization of the stent lumen and consequently determination of stent patency by cMRA.
RATIONALE AND OBJECTIVES: The objective of this study was to investigate the potential for artifact-free coronary magnetic resonance angiography (cMRA) in the presence of dedicated metallic MR stents in vitro and in a swine model. METHODS: All investigations were performed at 1.5 T, applying a standard cMRA gradient echo sequence with a T2 preparation pulse. Two prototypes of each hand-woven, mechanically woven, and lasered Aachen Resonance Coronary MR Stents made out of an MR-compatible metallic alloy and dilated to 2.5 mm and 4 mm were examined in a water bath. RESULTS: Artifact behavior was judged independently by 2 radiologists as showing "no artifacts" for all tested stent types. Signal-to-noise ratios inside and outside of the stents were measured yielding a Pearson correlation coefficient of 0.98 (y = 1.22 + 0.92x). Nineteen stents (8 hand woven, 3 mechanically woven, 8 lasered) were deployed in coronary arteries of 19 domestic pigs and were examined by cMRA. Artifact behavior of the stents was analyzed by measuring the signal-to-noise ratio at the stent positions and compared with signal-to-noise ratio measurements outside of the stents, yielding a Pearson correlation coefficient of 0.90 (y = -0.75 + 1.06x). CONCLUSIONS: All 3 prototypes of coronary MR stents allowed complete visualization of the stent lumen and consequently determination of stent patency by cMRA.
Authors: S C Gerretsen; M E Kooi; S Schalla; T Delhaas; G Snoep; J M A Van Engelshoven; T Leiner Journal: Cardiovasc J Afr Date: 2007 Jul-Aug Impact factor: 1.167