Measurement of the source size of a 6- and 18-MV radiotherapy linac.

E Loewenthal1, E Loewinger, E Bar-Avraham, G Barnea.   

Abstract

The quality of portal imaging is strongly affected by the source size of the radiotherapy machine. The effective source size of the dual-energy clinac 1800 (6 and 18 MV) was measured with a 50 microns wide and 120 mm long slit formed by two tungsten-copper alloy blocks. A series of slit images were obtained by translating the slit horizontally. The images were analyzed using a microdensitometer. The measured data was simulated using an analytical model of the source and its size was derived by a best-fit analysis. For both energies the FWHM was found to be 1.5 +/- 0.1 mm.

Entities:  

Mesh:

Year:  1992        PMID: 1508108     DOI: 10.1118/1.596812

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Monte Carlo Simulation of Siemens ONCOR Linear Accelerator with BEAMnrc and DOSXYZnrc Code.

Authors:  Keyvan Jabbari; Hossein Saberi Anvar; Mohammad Bagher Tavakoli; Alireza Amouheidari
Journal:  J Med Signals Sens       Date:  2013-07

2.  Modification and validation of an analytical source model for external beam radiotherapy Monte Carlo dose calculations.

Authors:  Scott E Davidson; Jing Cui; Stephen Kry; Joseph O Deasy; Geoffrey S Ibbott; Milos Vicic; R Allen White; David S Followill
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

3.  Characterization of linear accelerator X-ray source size using a laminated beam-spot camera.

Authors:  Collins Yeboah
Journal:  J Appl Clin Med Phys       Date:  2011-05-10       Impact factor: 2.102

4.  Characterization of linear accelerator X-ray source size using a laminated beam-spot camera.

Authors:  Collins Yeboah
Journal:  J Appl Clin Med Phys       Date:  2011-05-10       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.