New onset of pacemaker dependency after permanent pacemaker implantation.

Toshihisa Nagatomo1, Haruhiko Abe, Kan Kikuchi, Yasuhide Nakashima.   

Abstract

Pacemaker dependency has dangerous consequences under conditions of electromagnetic interference, unrecognized lead dysfunction, and battery depletion, and has been associated with cardiovascular and overall mortality. The aim of this study was to examine the incidence of new onset of pacemaker dependency during long-term follow-up after pacing system implantation. The study included 518 patients (mean age 72.9 +/- 10.4 years) who presented with intrinsic rhythms at the time of implantation. Indications for pacing were sick sinus syndrome (SSS) in 275 (53%) patients, AV block in 209 (40%), and AF with bradycardia in 34 (7%) patients. The mean follow-up was 3.7 +/- 2.7 years (range 1-17). Pacemaker dependency was defined as the absence of an intrinsic rhythm during backup pacing at 30 beats/min for 30 seconds. New onset of pacemaker dependency was observed in 23 (4.4%) of the 518 patients at a mean of 3.1 +/- 2.7 years of follow-up. Pacing indications were SSS in 6 (2.2%) of 275 patients, AV block in 15 (7.2%) of 209, and AF with bradycardia in 2 (5.9%) of 34. Patients with AV block had a significantly higher incidence of pacemaker dependency than patients with SSS (odds ratio 3.51; 95% CI 1.3 to 9.3; P = 0.012). The average annual rate of new pacemaker dependency was 1.6% during an 8-year follow-up. The incidence of new onset of pacemaker dependency varied among pacing indications, and was significantly higher in patients with AV block than patients with SSS.

Entities:  

Mesh:

Year:  2004        PMID: 15078400     DOI: 10.1111/j.1540-8159.2004.00466.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  6 in total

1.  eComment. Persistent annual risk for pacemaker implantation after aortic valve replacement.

Authors:  Jamil Hajj-Chahine
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-04

2.  Predictors of right ventricular pacing and pacemaker dependence in transcatheter aortic valve replacement patients.

Authors:  Esseim Sharma; Antony F Chu
Journal:  J Interv Card Electrophysiol       Date:  2017-12-19       Impact factor: 1.900

3.  Short-term pacemaker dependency after transcatheter aortic valve implantation.

Authors:  Christiana Schernthaner; Johannes Kraus; Franz Danmayr; Matthias Hammerer; Jens Schneider; Uta C Hoppe; Bernhard Strohmer
Journal:  Wien Klin Wochenschr       Date:  2016-01-08       Impact factor: 1.704

4.  Aortic Valve Calcification as a Predictor of Post-Transcatheter Aortic Valve Replacement Pacemaker Dependence.

Authors:  Esseim Sharma; Brian McCauley; Dhairyasheel S Ghosalkar; Michael Atalay; Scott Collins; Anshul Parulkar; Wasiq Sheikh; Malik B Ahmed; Antony Chu
Journal:  Cardiol Res       Date:  2020-05-03

5.  Long-term Intrinsic Rhythm Evaluation in Dogs with Atrioventricular Block.

Authors:  R A Santilli; D M Porteiro Vázquez; T Vezzosi; M Perego
Journal:  J Vet Intern Med       Date:  2015-11-17       Impact factor: 3.333

6.  Stability of intrinsic rhythm in pacemaker-dependent patients during pacemaker replacement: Can we predict the need for temporary pacing?

Authors:  Yuki Kimura; Masataka Sumiyoshi; Kenji Inoue; Masayuki Shiozaki; Kentaro Fukuda; Yasumasa Fujiwara; Haruna Tabuchi; Hidemori Hayashi; Gaku Sekita; Takashi Tokano; Yuji Nakazato; Hiroyuki Daida
Journal:  J Arrhythm       Date:  2018-05-28
  6 in total