Cross-calibration of ionization chambers in proton and carbon beams.

Tatsuaki Kanai1, Akifumi Fukumura, Yohsuke Kusano, Munefumi Shimbo, Teiji Nishio.   

Abstract

The calibration coefficients of a parallel plate ionization chamber are examined by comparing the coefficients obtained through three methods: a calculation from a 60Co calibration coefficient, N(D, omega, 60Co), a cross-calibration of a parallel plate ionization chamber using a cylindrical ionization chamber at the plateau region of a mono-energetic beam and a cross-calibration of the chamber using a cylindrical chamber at the middle of the SOBP of the therapeutic beams. This paper also examines reference conditions for determining absorbed dose to water in the cases of therapeutic carbon and proton beams. In the dose calibration procedure recommended by IAEA, irradiation fields should be larger than 10 cm in diameter and the water phantom should extend by at least 5 cm beyond each side of the field. These recommendations are experimentally verified for proton and carbon beams. For proton beams, the calibration coefficients obtained by these three methods approximately agreed. For carbon beams, the calibration coefficients obtained by the second method were about 1.0% larger than those obtained by the third method, and the calibration coefficients obtained by cross-calibration using 290 MeV/u beams were 0.5% lower than those obtained using 400 MeV/u beams. The calibration coefficient obtained by the first method agreed roughly with the results obtained by SOBP beams.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15070201     DOI: 10.1088/0031-9155/49/5/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Radiological properties of nanocomposite Fricke gel dosimeters for heavy ion beams.

Authors:  Takuya Maeyama; Nobuhisa Fukunishi; Kenichi L Ishikawa; Kazuaki Fukasaku; Shigekazu Fukuda
Journal:  J Radiat Res       Date:  2016-03-10       Impact factor: 2.724

2.  Contribution to dose in healthy tissue from secondary target fragments in therapeutic proton, He and C beams measured with CR-39 plastic nuclear track detectors.

Authors:  Satoshi Kodaira; Hisashi Kitamura; Mieko Kurano; Hajime Kawashima; Eric R Benton
Journal:  Sci Rep       Date:  2019-03-06       Impact factor: 4.379

3.  Whole Three-Dimensional Dosimetry of Carbon Ion Beams with an MRI-Based Nanocomposite Fricke Gel Dosimeter Using Rapid T1 Mapping Method.

Authors:  Shinya Mizukami; Yusuke Watanabe; Takahiro Mizoguchi; Tsutomu Gomi; Hidetake Hara; Hideyuki Takei; Nobuhisa Fukunishi; Kenichi L Ishikawa; Shigekazu Fukuda; Takuya Maeyama
Journal:  Gels       Date:  2021-11-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.