Cesare Sabbioni1, Maria Addolorata Saracino, Roberto Mandrioli, Sergio Pinzauti, Sandra Furlanetto, Gilberto Gerra, Maria Augusta Raggi.
Abstract
The comparison of two HPLC methods, one with electrochemical detection and the other with coulometric detection, for the simultaneous analysis of catecholamines and 4-hydroxy-3-methoxyphenylethylene glycol (MHPG) in human plasma is presented. The careful pre-treatment of plasma samples is based on an innovative two-step procedure by means of solid-phase extraction (SPE) which uses one single hydrophilic-lipophilic balance cartridge. The extraction yield values found were higher than 85% for epinephrine, norepinephrine and MHPG, and higher than 70% for dopamine. The assays carried out on real plasma samples with the coulometric system gave good results in terms of sensitivity (limits of quantitation: 0.10-0.15 ng ml(-1) for catecholamines, 0.6 ng ml(-1) for MHPG) and selectivity, while interference was sometimes found when using the amperometric system. Precision was also satisfactory, with relative standard deviation values for intermediate precision always lower than 6%. The HPLC method with coulometric detection coupled to a novel SPE procedure is thus suitable for the simultaneous determination of catecholamines and MHPG in plasma of volunteers subjected to experimental stress.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15065778 DOI: 10.1016/j.chroma.2004.01.008