Measurements of the dosimetric constants for a new 103Pd brachytherapy source.

Stephen W Peterson1, Bruce Thomadsen.   

Abstract

PURPOSE: This report provides dosimetry information for a new brachytherapy 103Pd source (Model 2335) manufactured by Best Medical International and, through comparisons with data from another published report for the same source, presents a suggested hybrid dataset for clinical applications. METHODS AND MATERIALS: Dose measurements were made using thermoluminescent dosimeters (TLD), with the 103Pd source in the center of a Virtual Water phantom on a rotating insert, allowing for a number of different possible angles, and 12 TLD cubes (1.0 x 1.0 x 1.0 mm) and 16 TLD rods (1.0 x 1.0 x 3.0 mm) arranged in an outward spiral pattern at distances ranging from 0.5 to 5.0 cm. All measurements are based on the 2000 correction of the National Institute of Standards and Technology 1999 standard.
RESULTS: From these measurements, tables are presented for the radial dose function and the anisotropy function. The dose rate constant=0.71 +/- 0.07 cGy h(-1) U(-1) and the anisotropy constant=0.96 +/- 0.03.
CONCLUSIONS: The dose rate constant for the hybrid=0.70 and the anisotropy constant for the hybrid=0.92, Hybrid tables are presented.

Mesh:

Substances:

Year:  2002        PMID: 15062179     DOI: 10.1016/s1538-4721(02)00015-6

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  3 in total

1.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

2.  Perturbation of TG-43 parameters of the brachytherapy sources under insufficient scattering materials.

Authors:  Mehdi Zehtabian; Sedigheh Sina; Reza Faghihi; Ali Meigooni
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

3.  Comparison of TG-43 dosimetric parameters of brachytherapy sources obtained by three different versions of MCNP codes.

Authors:  Neda Zaker; Mehdi Zehtabian; Sedigheh Sina; Craig Koontz; Ali S Meigooni
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.