Literature DB >> 1505487

Technical failure of biliary metal stent deployment in a series of 116 applications.

N Bethge1, H J Wagner, K Knyrim, H B Zimmermann, E Starck, J Pausch, N Vakil.   

Abstract

Biliary metal stents are thought to offer improved long-term palliation of malignant biliary obstruction due to a lower incidence of migration and clogging. Placement of these stents is technically more complicated than that of plastic endoprostheses and requires two experienced physicians. We report the incidence and reasons for apparent malfunction of expandable metal stent deployment (Wallstents and Strecker stents). In 116 applications of 82 Wallstents (endoscopic approach: n = 33, transhepatic approach: n = 49), we observed 19 cases of stent malfunction due to technical problems of stent delivery. In 13 cases (15.8%), the restraining membrane of the Wallstent could not be retracted sufficiently to deliver the stent. There were 6 (17.6%) failures in 34 cases of Strecker stent deployment. In 3 cases, we noted difficult balloon removal, including avulsion of the balloon catheter shaft within the endoscope during attempted balloon removal in one case. In one case, the Strecker stent could only be released partially, requiring subsequent endoscopic extraction. In two patients, only partial expansion of one end of the Strecker stent could be achieved. Given the significant malfunction rate of expandable metal stents during stent delivery, further improvements in the delivery system of the metal stents are required.

Entities:  

Mesh:

Year:  1992        PMID: 1505487     DOI: 10.1055/s-2007-1010506

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  3 in total

Review 1.  Endoscopic stenting for palliation of malignant biliary obstruction. A review of progress in the last 15 years.

Authors:  J J Sung; S C Chung
Journal:  Dig Dis Sci       Date:  1995-06       Impact factor: 3.199

2.  A three year follow up of self expanding metal stents in the endoscopic palliation of longterm survivors with malignant biliary obstruction.

Authors:  S O'Brien; A R Hatfield; P I Craig; S P Williams
Journal:  Gut       Date:  1995-04       Impact factor: 23.059

3.  A comparative study on the efficacy of covered metal stent and plastic stent in unresectable malignant biliary obstruction.

Authors:  Jae Myoung Choi; Jin Hong Kim; Soon Sun Kim; Jun Hwan Yu; Jae Chul Hwang; Byung Moo Yoo; Sang Heum Park; Ho Gak Kim; Dong Ki Lee; Kang Hyun Ko; Kyo Sang Yoo; Do Hyun Park
Journal:  Clin Endosc       Date:  2012-03-31
  3 in total