Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator.

Etsuo Kunieda1, Osamu Kawaguchi, Hidetoshi Saitoh, Tatsuya Fujisaki, Atsuya Takeda, Takatsugu Kawase, Hossain M Deloar, Naoyuki Shigematsu, Atsushi Kubo.   

Abstract

We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41+/-0.25 mm.