| Literature DB >> 15036851 |
Etsuo Kunieda1, Osamu Kawaguchi, Hidetoshi Saitoh, Tatsuya Fujisaki, Atsuya Takeda, Takatsugu Kawase, Hossain M Deloar, Naoyuki Shigematsu, Atsushi Kubo.
Abstract
We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41+/-0.25 mm.Mesh:
Year: 2004 PMID: 15036851 DOI: 10.1016/j.radonc.2003.11.008
Source DB: PubMed Journal: Radiother Oncol ISSN: 0167-8140 Impact factor: 6.280