Literature DB >> 15030842

Folding procedure for acrylic intraocular lenses.

Rita Mencucci1, Rosanna Dei, Daniele Danielli, Massimo Susini, Ugo Menchini.   

Abstract

PURPOSE: To compare in vitro the effect of 2 standard methods of folding acrylic intraocular lenses (IOLs) on surface characteristics and bacterial adhesion.
SETTING: Eye Clinic and Department of Health-Microbiology Unit, University of Florence, Florence, Italy.
METHODS: To evaluate the effect of folding, 2 types of acrylic IOLs were not folded or folded with a forceps or an injector and then processed for scanning electron microscopy (SEM) examination. Bacterial adhesion was assessed using an ocular isolate of Pseudomonas aeruginosa. Nonfolded and folded IOLs were placed in test tubes containing the bacterial suspension for direct counting of viable adherent bacteria and for SEM.
RESULTS: The injector-folded IOLs did not show major alterations on the surface; 5 of the 9 forceps-folded IOLs showed marks or scratches in the profile of the optic. The mean number of viable adherent bacteria per area of IOL optic was 1082 (95% confidence interval [CI], 835-1330) in forceps-folded IOLs, 366 (95% CI, 192-359) in injector-folded IOLs, and 206 (95% CI, 123-289) in nonfolded IOLs. Scanning electron microscopy confirmed more surface irregularities on forceps-folded IOLs, with bacteria adherent preferentially on the surface scratches.
CONCLUSION: Forceps-folding provoked more surface irregularities, which probably make IOLs more susceptible to bacterial adhesion.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15030842     DOI: 10.1016/j.jcrs.2003.11.025

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Clinical outcomes of combined pars plana vitrectomy and trans-scleral 4-point suture fixation of a foldable intraocular lens.

Authors:  Neha Goel
Journal:  Eye (Lond)       Date:  2018-02-05       Impact factor: 3.775

2.  Innovative intraocular lens loading magnified.

Authors:  Rengaraj Venkatesh; Bharat Gurnani; Kirandeep Kaur
Journal:  Indian J Ophthalmol       Date:  2020-08       Impact factor: 1.848

3.  Evaluation of semi-preloaded intraocular lens delivery system.

Authors:  Rajesh Rajesh
Journal:  Rom J Ophthalmol       Date:  2019 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.