Literature DB >> 15022249

Experimental application of pulsed Ho:YAG laser-induced liquid jet as a novel rigid neuroendoscopic dissection device.

Tomohiro Ohki1, Atsuhiro Nakagawa, Takayuki Hirano, Tokitada Hashimoto, Viren Menezes, Hidefumi Jokura, Hiroshi Uenohara, Yasuhiko Sato, Tsutomu Saito, Reizo Shirane, Teiji Tominaga, Kazuyoshi Takayama.   

Abstract

BACKGROUND AND OBJECTIVES: Although water jet technology has been considered as a feasible neuroendoscopic dissection methodology because of its ability to perform selective tissue dissection without thermal damage, problems associated with continuous use of water and the ensuing fountain-effect-with catapulting of the tissue-could make water jets unsuitable for endoscopic use, in terms of safety and ease of handling. Therefore, the authors experimented with minimization of water usage during the application of a pulsed holmium:yttrium-aluminum-garnet (Ho:YAG) laser-induced liquid jet (LILJ), while assuring the dissection quality and the controllability of a conventional water jet dissection device. We have developed the LILJ generator for use as a rigid neuroendoscope, discerned its mechanical behavior, and evaluated its dissection ability using the cadaveric rabbit ventricular wall. STUDY DESIGN/
MATERIALS AND METHODS: The LILJ generator is incorporated into the tip of a stainless steel tube (length: 22 cm; internal diameter: 1.0 mm; external diameter: 1.4 mm), so that the device can be inserted into a commercial, rigid neuroendoscope. Briefly, the LILJ is generated by irradiating an internally supplied water column within the stainless steel tube using the pulsed Ho:YAG laser (wave length: 2.1 microm, pulse duration time: 350 microseconds) and is then ejected through the metal nozzle (internal diameter: 100 microm). The Ho:YAG laser pulse energy is conveyed through optical quartz fiber (core diameter: 400 microm), while cold water (5 degrees C) is internally supplied at a rate of 40 ml/hour. The relationship between laser energy (range: 40-433 mJ/pulse), standoff distance (defined as the distance between the tip of the optical fiber and the nozzle end; range: 10-30 mm), and the velocity, shape, pressure, and average volume of the ejected jet were analyzed by means of high-speed camera, PVDF needle hydrophone, and digital scale. The quality of the dissection plane, the preservation of blood vessels, and the penetration depth were evaluated using five fresh cadaveric rabbit ventricular walls, under neuroendoscopic vision.
RESULTS: Jet velocity (7.0-19.6 m/second) and pressure (0.07-0.28 MPa) could be controlled by varying the laser energy, which determined the penetration depth in the cadaveric rabbit ventricular wall (0.07-1.30 mm/shot). The latter could be cut into desirable shapes-without thermal effects-under clear neuroendoscopic vision. The average volume of a single ejected jet could be confined to 0.42-1.52 microl/shot, and there was no accompanying generation of shock waves. Histological specimens revealed a sharp dissection plane and demonstrated that blood vessels of diameter over 100 microm could be preserved, without thermal damage.
CONCLUSIONS: The present pulsed LILJ system holds promise as a safe and reliable dissection device for deployment in a rigid neuroendoscope. Copyright 2004 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2004        PMID: 15022249     DOI: 10.1002/lsm.20021

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  6 in total

1.  A laser-induced pulsed water jet for layer-selective submucosal dissection of the esophagus.

Authors:  T Nakano; C Sato; M Yamada; A Nakagawa; H Yamamoto; F Fujishima; T Tominaga; S Satomi; N Ohuchi
Journal:  Laser Ther       Date:  2016-10-01

2.  Application of actuator-driven pulsed water jet for coronary artery bypass grafting: assessment in a swine model.

Authors:  Tomoyuki Suzuki; Shunsuke Kawamoto; Atsuhiro Nakagawa; Toshiki Endo; Teiji Tominaga; Masatoshi Akiyama; Osamu Adachi; Kiichiro Kumagai; Yoshikatsu Saiki
Journal:  J Artif Organs       Date:  2017-11-16       Impact factor: 1.731

Review 3.  Use of water jet instruments in gastrointestinal endoscopy.

Authors:  Toru Nakano; Chiaki Sato; Tadashi Sakurai; Takashi Kamei; Atsuhiro Nakagawa; Noriaki Ohuchi
Journal:  World J Gastrointest Endosc       Date:  2016-02-10

4.  Experimental Application of Piezoelectric Actuator-Driven Pulsed Water Jets in Retinal Vascular Surgery.

Authors:  Hiroshi Kunikata; Yuji Tanaka; Naoko Aizawa; Atsuhiro Nakagawa; Teiji Tominaga; Toru Nakazawa
Journal:  Transl Vis Sci Technol       Date:  2014-12-17       Impact factor: 3.283

5.  Evaluation of a newly developed piezo actuator-driven pulsed water jet system for liver resection in a surviving swine animal model.

Authors:  Chikashi Nakanishi; Toru Nakano; Atsuhiro Nakagawa; Chiaki Sato; Masato Yamada; Naoki Kawagishi; Teiji Tominaga; Noriaki Ohuchi
Journal:  Biomed Eng Online       Date:  2016-01-25       Impact factor: 2.819

6.  Repetitive regime of highly focused liquid microjets for needle-free injection.

Authors:  Jan Krizek; Paul Delrot; Christophe Moser
Journal:  Sci Rep       Date:  2020-03-19       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.