The Accoson Greenlight 300, the first non-automated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults.

John W Graves1, Mary Tibor, Blaithnead Murtagh, Lois Klein, Sheldon G Sheps.   

Abstract

BACKGROUND: Accurate blood pressure measurement is critical to successful care of the hypertensive patient. The mercury manometer is rapidly being removed from clinical practice due to concerns about environmental contamination. The Accoson Greenlight 300 device is the first non-automated mercury-free device for auscultatory blood pressure measurement to pass the International Protocol for validation of blood pressure measuring devices in adults.
METHODS: Fifty-one patients were enrolled to obtain the 33 patients required to fill the three systolic blood pressure (SBP) ranges (low 90-129 mmHg, medium 130-160 mmHg, high 161-180 mmHg) and diastolic blood pressure (DBP) ranges (low 40-79 mmHg, medium 80-100 mmHg, high 101-130 mmHg). Nine sequential blood pressure measurements by two observers with the mercury manometer and the supervisor with the Greenlight 300 were taken. The first two observer blood pressures were used to classify the SBP and DBP range. The readings were then analyzed in two phases to determine whether the device passed the International Protocol.
RESULTS: The device passed Phase 1 using 15 subjects. In Phase 2.1 (n=33) for the 99 readings of SBP, 84 of 99 were within 5 mmHg, 95 of 99 were within 10 mmHg, and 98 of 99 were within 15 mmHg. For the 99 readings of DBP, 74 of 99 were within 5 mmHg, 90 of 99 were within 10 mmHg, and 96 of 99 were within 15 mmHg. In Phase 2.2 (n=33) for SBP, 33 of 33 had 2 of 3 SBP within 5 mmHg and none had all three readings >or=5 mmHg. For DBP, 27 of 33 had 2 of 3 of their DBP within 5 mmHg and only three had all three readings > or =5 mmHg.
CONCLUSIONS: The Accoson Greenlight 300 is the first electronic non-automated blood pressure measurement device to pass the International Protocol.

Entities:  

Mesh:

Year:  2004        PMID: 15021073     DOI: 10.1097/00126097-200402000-00004

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  8 in total

1.  Estimation of mean arterial pressure from the oscillometric cuff pressure: comparison of different techniques.

Authors:  Dingchang Zheng; John N Amoore; Stephan Mieke; Alan Murray
Journal:  Med Biol Eng Comput       Date:  2010-11-02       Impact factor: 2.602

Review 2.  Blood pressure measurement: clinic, home, ambulatory, and beyond.

Authors:  Paul E Drawz; Mohamed Abdalla; Mahboob Rahman
Journal:  Am J Kidney Dis       Date:  2012-04-21       Impact factor: 8.860

3.  Blood pressure-lowering efficacy of the fixed-dose combination of azilsartan medoxomil and chlorthalidone: a factorial study.

Authors:  Domenic Sica; George L Bakris; William B White; Michael A Weber; William C Cushman; Patrick Huang; Andrew Roberts; Stuart Kupfer
Journal:  J Clin Hypertens (Greenwich)       Date:  2012-03-06       Impact factor: 3.738

4.  Recommendations for blood pressure measurement in humans: an AHA scientific statement from the Council on High Blood Pressure Research Professional and Public Education Subcommittee.

Authors:  Thomas G Pickering; John E Hall; Lawrence J Appel; Bonita E Falkner; John W Graves; Martha N Hill; Daniel H Jones; Theodore Kurtz; Sheldon G Sheps; Edward J Roccella
Journal:  J Clin Hypertens (Greenwich)       Date:  2005-02       Impact factor: 3.738

5.  Comparison of aneroid and oscillometric blood pressure measurements in children.

Authors:  Sigridur B Eliasdottir; Sandra D Steinthorsdottir; Olafur S Indridason; Runolfur Palsson; Vidar O Edvardsson
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-09-20       Impact factor: 3.738

6.  Comparison of the accuracy and errors of blood pressure measured by 2 types of non-mercury sphygmomanometers in an epidemiological survey.

Authors:  SeongIl Choi; Yu-Mi Kim; Jinho Shin; Young-Hyo Lim; Sung-Yong Choi; Bo-Youl Choi; Kyung-Won Oh; Hyung-Min Lee; Kyung-Ji Woo
Journal:  Medicine (Baltimore)       Date:  2018-06       Impact factor: 1.889

7.  Blood pressure measurements and hypertension in infants, children, and adolescents: from the postmercury to mobile devices.

Authors:  Seon Hee Lim; Seong Heon Kim
Journal:  Clin Exp Pediatr       Date:  2021-09-15

8.  A randomized titrate-to-target study comparing fixed-dose combinations of azilsartan medoxomil and chlorthalidone with olmesartan and hydrochlorothiazide in stage-2 systolic hypertension.

Authors:  William C Cushman; George L Bakris; William B White; Michael A Weber; Domenic Sica; Andrew Roberts; Eric Lloyd; Stuart Kupfer
Journal:  J Hypertens       Date:  2018-04       Impact factor: 4.844

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.