Literature DB >> 15010384

Assessment of a technique for 2D-3D registration of cerebral intra-arterial angiography.

J V Byrne1, C Colominas, J Hipwell, T Cox, J A Noble, G P Penney, D J Hawkes.   

Abstract

This study assesses the ability of a computer algorithm to perform automated 2D-3D registrations of digitally subtracted cerebral angiograms. The technique was tested on clinical studies of five patients with intracranial aneurysms. The automated procedure was compared against a gold standard manual registration, and achieved a mean registration accuracy of 1.3 mm (SD 0.6 mm). Two registration strategies were tested using coarse (128 x 128 pixel) or fine (256 x 256 pixel) images. The mean registration errors proved similar but registration of the lower resolution images was 3 times quicker (mean registration times 33 s, SD 13 s for low and 150 s SD 48 s for high resolution images). The automated techniques were considerably faster than manual registrations but achieved similar accuracy. The technique has several potential uses but is particularly applicable to endovascular treatment techniques.

Entities:  

Mesh:

Year:  2004        PMID: 15010384     DOI: 10.1259/bjr/27339681

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  1 in total

1.  A comparative study on manual and automatic slice-to-volume registration of CT images.

Authors:  Laura Frühwald; Joachim Kettenbach; Michael Figl; Johann Hummel; Helmar Bergmann; Wolfgang Birkfellner
Journal:  Eur Radiol       Date:  2009-06-06       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.