Accuracy of the Justy II Apex locator in determining working length in simulated horizontal and vertical fractures.

M Azabal1, D Garcia-Otero, J C de la Macorra.   

Abstract

AIM: To study the effectiveness of an electronic apex locator (EAL; Justy II; Yoshida Dentcraft, Tokyo, Japan) in locating simulated horizontal and vertical fractures in single roots.
METHODOLOGY: An EAL was used to measure the distance within the canal of horizontal (n = 31) and vertical (n = 31) fractures, created with a disk in single-rooted teeth. Accuracy of the EAL was evaluated by comparing the measurements with those made using a size 10 file. Data were analysed with the non-parametric Passing and Bablok method.
RESULTS: For simulated horizontal fractures, the EAL measured exactly the same length as a size 10 file, without constant or proportional errors. In vertical simulated fractures, the EAL measured (on average) with a constant error of 7.5 mm shorter than the size 10 file; the difference had a wide confidence interval (-72.3 to 2.6 mm).
CONCLUSIONS: In this laboratory study, the Justy II EAL was able to determine accurately the position of simulated horizontal fractures, but was unreliable when measuring simulated vertical fractures.

Entities:  

Mesh:

Year:  2004        PMID: 15009406     DOI: 10.1111/j.0143-2885.2004.00776.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  2 in total

1.  Precision of multi-frequency electronic apex locators.

Authors:  Roy George
Journal:  Evid Based Dent       Date:  2016-09

2.  In vitro evaluation of the accuracy of two electronic apex locators.

Authors:  Viresh Chopra; Shibani Grover; S Datta Prasad
Journal:  J Conserv Dent       Date:  2008-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.