The validity of emergency department triage blood pressure measurements.

John J Cienki1, Lawrence A DeLuca, Natalie Daniel.   

Abstract

OBJECTIVE: Automated blood pressure (ABP) devices are ubiquitous at emergency department (ED) triage. Previous studies failed to evaluate ABP devices against accepted reference standards or demonstrate triage readings as accurate reflections of blood pressure (BP). This study evaluated ED triage measurements made using an ABP device and assessed agreement between triage BP and BP taken under recommended conditions.
METHODS: A prospective study was conducted at an urban teaching hospital. Patients were enrolled by convenience sampling. Simultaneous automated and manual triage BPs were obtained using one BP cuff with a Y-tube connector. Research assistants were certified in obtaining manual BP as described by the British Hypertension Society (BHS). Patients were placed in a quiet setting, and manual BP was repeated by American Heart Association (AHA) standards. Data analysis was performed using methods described by Bland and Altman. The ABP device was assessed using Association for the Advancement of Medical Instrumentation (AAMI) and BHS criteria.
RESULTS: One hundred seventy-one patients were enrolled. Systolic BP (sBP) range was 81 to 218 mm Hg; diastolic BP (dBP) range was 43 to 130 mm Hg. Automated vs. manual sBP difference was 3.8 +/- 11.2 mm Hg (95% confidence interval [CI] = 2.1 to 5.4); dBP difference was 6.6 +/- 9.0 mm Hg (95% CI = -7.9 to -5.2). Manual triage BP vs. AHA standard SBP difference was 11.6 +/- 12.8 mm Hg (95% CI = 9.1 to 14.1); dBP difference was 9.9 +/- 10.4 mm Hg (95% CI = 7.9 to 12.0). The ABP device failed to meet AAMI criteria and received a BHS rating of "D." Poor operator technique and extraneous patient and operator movement appeared to hamper accuracy.
CONCLUSIONS: ABP triage measurements show significant discrepancies from a reference standard. Repeat measurements following AHA standards demonstrate significant decreases in the measured blood pressures.

Entities:  

Mesh:

Year:  2004        PMID: 15001402     DOI: 10.1111/j.1553-2712.2004.tb02203.x

Source DB:  PubMed          Journal:  Acad Emerg Med        ISSN: 1069-6563            Impact factor:   3.451


  6 in total

1.  Out-of-hospital characteristics and care of patients with severe sepsis: a cohort study.

Authors:  Christopher W Seymour; Roger A Band; Colin R Cooke; Mark E Mikkelsen; Julie Hylton; Tom D Rea; Christopher H Goss; David F Gaieski
Journal:  J Crit Care       Date:  2010-04-08       Impact factor: 3.425

Review 2.  Hypertension in the Emergency Department.

Authors:  Stewart Siu-Wa Chan; Colin A Graham; T H Rainer
Journal:  Curr Hypertens Rep       Date:  2016-04       Impact factor: 5.369

3.  Course of untreated high blood pressure in the emergency department.

Authors:  John J Cienki; Lawrence A Deluca; Daniel J Feaster
Journal:  West J Emerg Med       Date:  2011-11

Review 4.  Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.

Authors:  Noa Kallioinen; Andrew Hill; Mark S Horswill; Helen E Ward; Marcus O Watson
Journal:  J Hypertens       Date:  2017-03       Impact factor: 4.844

5.  Association of Elevated Blood Pressure in the Emergency Department With Chronically Elevated Blood Pressure.

Authors:  Sabrina J Poon; Christianne L Roumie; Colin J O'Shea; Daniel Fabbri; Joseph R Coco; Sean P Collins; Phillip D Levy; Candace D McNaughton
Journal:  J Am Heart Assoc       Date:  2020-06-06       Impact factor: 5.501

6.  Design and validation of a new scale for prehospital evaluation of stroke and large vessel occlusion.

Authors:  Yanqi Shao; Zheyu Zhang; Bo Jin; Jingsi Xu; Deqing Peng; Yu Geng; Jungen Zhang; Sheng Zhang
Journal:  Ther Adv Neurol Disord       Date:  2022-06-30       Impact factor: 6.430

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.