Literature DB >> 14987078

Direct chemical analysis of solids by laser ablation in an ion storage time-of-flight mass spectrometer.

Gregory L Klunder1, Patrick M Grant, Brian D Andresen, Richard E Russo.   

Abstract

A laser ablation/ionization mass spectrometer system is described for the direct analysis of solids, particles, and fibers. The system uses a quadrupole ion trap operated in an ion storage mode, coupled with a reflectron time-of-flight mass spectrometer). The sample is inserted radially into the ring electrode, and an imaging system allows direct viewing and selected analysis of the sample. Measurements identified trace contaminants of Ag, Sn, and Sb in a Pb target with single laser shot experiments. Resolution (m/Delta m) of 1500 and detection limits of approximately 10 pg have been achieved with a single laser pulse. The system configuration and related operating principles for accurately measuring low concentrations of isotopes are described.

Entities:  

Year:  2004        PMID: 14987078     DOI: 10.1021/ac0303261

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  1 in total

1.  High irradiance laser ionization mass spectrometry for direct speciation of iron oxides.

Authors:  Bin Yan; Lingfeng Li; Quan Yu; Wei Hang; Jian He; Benli Huang
Journal:  J Am Soc Mass Spectrom       Date:  2010-03-31       Impact factor: 3.109