Literature DB >> 14982983

Variability of plasma anti-Xa activities with different lots of enoxaparin.

Robert C Gosselin1, Jeffery H King, Kim A Janatpour, William E Dager, Edward C Larkin, John T Owings.   

Abstract

BACKGROUND: Previous studies have indicated the variability of anti-Xa activity in different sources of heparin and the variability of different methods used for measuring anti-Xa activity. Manufacturers of low-molecular-weight heparins (LMWHs) determine each lot's anti-Xa activity against the World Health Organization standard, but little information is known about anti-Xa activity variation between lots of LMWH and the impact on reported anti-Xa activity in patient samples.
OBJECTIVE: To determine the variation of plasma anti-Xa activity in patients receiving enoxaparin when different lots of enoxaparin are used for test calibration.
METHODS: We obtained 7 lots of enoxaparin containing approximately 10,000 IU/mL and one lot containing approximately 15,000 IU/mL of anti-Xa activity. For each lot, a 2.0 anti-Xa IU/mL dilution was prepared and a calibration curve performed using a chromogenic method. To test the variation in reported results between the different calibration lots, 20 patient samples were tested. Nineteen patients receiving enoxaparin and one patient not receiving enoxaparin (negative control) were tested in a blinded fashion, and the changes in light absorbance recorded. Anti-Xa activity results from tested plasmas were then extrapolated from each enoxaparin lot calibration curve.
RESULTS: Using Student's paired t-test, there were statistically significant differences between the plasma anti-Xa activities generated from the various enoxaparin lots. In the range of 0.5-1.0 IU/mL of anti-Xa activity, 3 (4.2%) samples had a >0.2 IU/mL difference (maximum difference 0.33 IU/mL) in anti-Xa activity between 2 lots of enoxaparin. For samples that had supratherapeutic anti-Xa activities (1.0-1.5 IU/mL anti-Xa activity), there was a wider variation (>0.2 IU/mL) in anti-Xa activity, which may have resulted in a dosing change.
CONCLUSIONS: The statistical differences in plasma anti-Xa activities noted between enoxaparin lots are not clinically significant. However, anti-Xa activities in the upper therapeutic and supratherapeutic ranges (>1.0 IU/mL of anti-Xa activity) resulted in a deviation of >0.3 IU/mL in reported anti-Xa activity, which may result in dosing changes.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14982983     DOI: 10.1345/aph.1D245

Source DB:  PubMed          Journal:  Ann Pharmacother        ISSN: 1060-0280            Impact factor:   3.154


  4 in total

1.  Toward development of a point-of-care assay of enoxaparin anticoagulant activity in whole blood.

Authors:  Mario A Inchiosa; Suryanarayana Pothula; Keshar Kubal; Vajubhai T Sanchala; Iris Navarro
Journal:  J Thromb Thrombolysis       Date:  2011-07       Impact factor: 2.300

2.  The use of a HEMOCHRON JR. HEMONOX point of care test in monitoring the anticoagulant effects of enoxaparin during interventional coronary procedures.

Authors:  Soumaya El Rouby; Marc Cohen; Andrea Gonzales; Debra Hoppensteadt; Ted Lee; Marcia L Zucker; Khaula Khalid; Frank M Laduca; Jawed Fareed
Journal:  J Thromb Thrombolysis       Date:  2006-04       Impact factor: 2.300

Review 3.  Measuring anti-factor xa activity to monitor low-molecular-weight heparin in obesity: a critical review.

Authors:  Gregory Egan; Mary H H Ensom
Journal:  Can J Hosp Pharm       Date:  2015 Jan-Feb

4.  Monitoring of anticoagulant therapy in heart disease: considerations for the current assays.

Authors:  Mohammadali Boroumand; Hamidreza Goodarzynejad
Journal:  J Tehran Heart Cent       Date:  2010-05-31
  4 in total