Literature DB >> 14979697

Syringe-cartridge solid-phase extraction method for patulin in apple juice.

Thomas A Eisele1, Midori Z Gibson.   

Abstract

A syringe-cartridge solid-phase extraction (SPE) method was developed for determination of patulin in apple juice. A 2.5 mL portion of test sample was passed through a conditioned macroporous SPE cartridge and washed with 2 mL 1% sodium bicarbonate followed by 2 mL 1% acetic acid. Patulin was eluted with 1 mL 10% ethyl acetate in ethyl ether and determined by reversed-phase liquid chromatography using a mobile phase consisting of 81% acetonitrile, 9% water, and 10% 0.05M potassium phosphate buffer, pH 2.4. Recoveries averaged 92% and the relative standard deviation was 8.0% in test samples spiked with 50 ng/mL patulin. The method appears to be applicable for monitoring apple juice samples to meet the U.S. Food and Drug Administration compliance action level of 50 microg/kg in an industrial quality assurance laboratory environment.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14979697

Source DB:  PubMed          Journal:  J AOAC Int        ISSN: 1060-3271            Impact factor:   1.913


  1 in total

1.  Determination of patulin by online-SPE-LC.

Authors:  A S Wendt; K P Raezke; P Winterhalter
Journal:  Mycotoxin Res       Date:  2006-12       Impact factor: 3.833