Literature DB >> 14975968

CT density measurements for characterization of adrenal tumors ex vivo: variability among three CT scanners.

Alfred Stadler1, Wolfgang Schima, Gerhard Prager, Peter Homolka, Gertraud Heinz, Sanjay Saini, Edith Eisenhuber, Bruno Niederle.   

Abstract

OBJECTIVE: Many studies have suggested that Hounsfield measurements on unenhanced CT can reliably differentiate adrenal adenomas from nonadenomas using a scanner-independent threshold level. The purpose of this study was to determine whether establishment of a scanner-independent threshold for differentiation of adenomas from nonadenomas is technically feasible.
MATERIALS AND METHODS: Surgically resected adrenal tumor specimens (total, seven; adenomas, three; nonadenomas, four; size range, 17-76 mm), were placed in an anthropomorphic phantom. Lesion specimens were scanned with one MDCT and two single-detector scanners. Scanning protocols for all three scanners included variations in kilovoltage (140, 120, and 80 [Somatom Plus 4, Somatom VolumeZoom] or 100 [Tomoscan AV] kVp) and slice thickness. Hounsfield measurements were performed on exactly matched slices using regions of interest of a constant size.
RESULTS: The difference in lesion Hounsfield measurements among scanning protocols with 140, 120, and 100/80 kVp was up to 6.2 H for the adenoma group and up to 3.8 H for the nonadenoma group. The comparison of the Tomoscan AV and the Somatom Plus 4 scanners showed a mean difference of 2.6 H at 120 kVp and of 4.6 H at 140 kVp. The differences between the Tomoscan AV and Somatom VolumeZoom scanners were 1.7 and 3.6 H for 120 and 140 kVp, respectively. Between the two Somatom scanners, the divergence was 2.9 and 3.3 H for the two kilovoltage settings. Differentiation between adenomas and nonadenomas was better at lower kilovoltage. Slice thickness did not affect the CT density measurements significantly.
CONCLUSION: Significant differences in CT density measurements of adrenal tumors may occur when different CT scanners or imaging protocols are used. The dependence of measurements on scanner type and scanning technique makes the recommendation of a universal, scanner- and protocol-independent threshold problematic.

Entities:  

Mesh:

Year:  2004        PMID: 14975968     DOI: 10.2214/ajr.182.3.1820671

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  6 in total

1.  Quantitative contrast-enhanced computed tomography: is there a need for system calibration?

Authors:  Kenneth A Miles; Helen Young; Sandra L Chica; Peter D Esser
Journal:  Eur Radiol       Date:  2006-09-29       Impact factor: 5.315

2.  Differentiation of neoplastic from bland macroscopic portal vein thrombi using dual-energy spectral CT imaging: a pilot study.

Authors:  Li Jun Qian; Jiong Zhu; Zhi Guo Zhuang; Qiang Xia; Yu Fan Cheng; Jian Ying Li; Jian Rong Xu
Journal:  Eur Radiol       Date:  2012-05-24       Impact factor: 5.315

3.  Utility of Intermediate-Delay Washout CT Images for Differentiation of Malignant and Benign Adrenal Lesions: A Multivariate Analysis.

Authors:  Chaan S Ng; Emre Altinmakas; Wei Wei; Payel Ghosh; Xiao Li; Elizabeth G Grubbs; Nancy D Perrier; Jeffrey E Lee; Victor G Prieto; Brian P Hobbs
Journal:  AJR Am J Roentgenol       Date:  2018-06-27       Impact factor: 3.959

4.  CT Hounsfield numbers of soft tissues on unenhanced abdominal CT scans: variability between two different manufacturers' MDCT scanners.

Authors:  Ramit Lamba; John P McGahan; Michael T Corwin; Chin-Shang Li; Tien Tran; J Anthony Seibert; John M Boone
Journal:  AJR Am J Roentgenol       Date:  2014-11       Impact factor: 3.959

5.  Myocardial density analysis utilizing automated myocardial defect analysis software on resting 320-detector MDCT.

Authors:  John M Troupis; Alex Karge; Sujith Seneviratne; Arthur Nasis; Eileen C Ang; Brian S Ko; Dee Nandurkar; Eldho Paul; Roland Hilling-Smith; James Cameron
Journal:  Int J Cardiovasc Imaging       Date:  2013-01-03       Impact factor: 2.357

6.  MDCT in the differentiation of adrenal masses: comparison between different scan delays for the evaluation of intralesional washout.

Authors:  Giuseppe Angelelli; Maria Elisabetta Mancini; Marco Moschetta; Pasquale Pedote; Pasquale Pignataro; Arnaldo Scardapane
Journal:  ScientificWorldJournal       Date:  2013-03-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.