Literature DB >> 1489865

Multicentre evaluation of the Boehringer Mannheim/Hitachi 747 analysis system.

P Bonini1, F Ceriotti, F Keller, P Brauer, H Stolz, C Pascual, L García Beltrán, D J Vonderschmitt, P Pei.   

Abstract

Analytical performance and practicability of the new Boehringer Mannheim/Hitachi 747 analysis system were assessed in a multicentre evaluation involving four laboratories. The analytical performance was evaluated according to a protocol similar to the ECCLS guidelines and comprised 13 analytes including enzymes, substrates and electrolytes. About 65,000 results were obtained within three months. The evaluation was planned and supported by a program system called "Computer Aided Evaluation". Acceptance criteria have been established for judging the results. The median of the within-run coefficients of variation (CVs) in control sera of all methods was below 1%, being far below the acceptance limit of 2%. The median of CVs of between-days imprecision was below 2% (acceptance criterion 3%). The high degree of precision prompted us to set up a biometrical model suitable for the differentiation between deviant points, outliers and measurements that can still be explained by the system performance. No relevant drift effects were observed during eight hours. The methods were linear over a wide range, avoiding rerun analysis in most cases. No sample-related carry-over was found. Reagent-dependent carry-over outside the acceptance limits was measured from uric acid to phosphorus to a slight extent, and from triacylglycerols to lipase, as well as from total protein to bilirubin to a perceptible degree. It can be avoided by separating these reagent combinations in the channel arrangement. Taking a systematic deviation of more than 10% as unacceptable, four of the 13 analytes suffered from interference by haemoglobin, one by bilirubin and one by turbidity. The Boehringer Mannheim/Hitachi 747 analysis system is capable of determining serum indices which in combination with the interferogram allow an assessment of the interference. With the exception of chloride the recovery of the assigned values for all control sera showed values between 95 and 105%. Out of 40 method comparison studies for enzymes and substrates, 31 yielded regression equations with less than 5% proportional errors and less than 5% constant errors. Deviations exceeding these acceptance criteria can be explained by differences in the reagent formulation, in the method employed or in calibration. The agreement of the ISE method comparisons was within a +/- 5% deviation over a wide analytical range. Practicability of the Boehringer Mannheim/Hitachi 747 analysis system was assessed with the help of a questionnaire, in which properties of the instrument were quantified, thus permitting a relatively objective rating. The 190 questions were placed in 14 groups, each dealing with an attribute of the instrument.(ABSTRACT TRUNCATED AT 400 WORDS)

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1489865

Source DB:  PubMed          Journal:  Eur J Clin Chem Clin Biochem        ISSN: 0939-4974


  4 in total

1.  Patient- and physician-related risk factors for hyperkalaemia in potassium-increasing drug-drug interactions.

Authors:  Emmanuel Eschmann; Patrick E Beeler; Vladimir Kaplan; Markus Schneemann; Gregor Zünd; Jürg Blaser
Journal:  Eur J Clin Pharmacol       Date:  2013-10-23       Impact factor: 2.953

2.  Acute psychological and physiological effects of psilocybin in healthy humans: a double-blind, placebo-controlled dose-effect study.

Authors:  Felix Hasler; Ulrike Grimberg; Marco A Benz; Theo Huber; Franz X Vollenweider
Journal:  Psychopharmacology (Berl)       Date:  2003-11-13       Impact factor: 4.530

3.  Sirolimus induced phosphaturia is not caused by inhibition of renal apical sodium phosphate cotransporters.

Authors:  Maria Haller; Stefan Amatschek; Julia Wilflingseder; Alexander Kainz; Bernd Bielesz; Ivana Pavik; Andreas Serra; Nilufar Mohebbi; Jürg Biber; Carsten A Wagner; Rainer Oberbauer
Journal:  PLoS One       Date:  2012-07-30       Impact factor: 3.240

4.  Increasing Efficiency and Quality by Consolidation of Clinical Chemistry and Immunochemistry Systems with MODULAR ANALYTICS SWA.

Authors:  Paolo Mocarelli; Gary L Horowitz; Pier Mario Gerthoux; Rossana Cecere; Roland Imdahl; Janneke Ruinemans-Koerts; Hilmar Luthe; Silvia Pesudo Calatayud; Marie Luisa Salve; Albert Kunst; Margaret McGovern; Katherine Ng; Wolfgang Stockmann
Journal:  J Autom Methods Manag Chem       Date:  2008
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.