Interface load analysis for computer-aided design of below-knee prosthetic sockets.

D P Reynolds1, M Lord.   

Abstract

A finite-element analysis is made for the compression of soft tissues of the residual lower limb contained in a prosthetic socket. The analysis is relevant to static loading during stance in a patellar-tendon-bearing, below-knee design of socket. Values of Young's modulus are obtained experimentally for use in the model. One of the main objectives is to study the sensitivity of the loading to these values and also to other assumed conditions. Using direct pressure at the limb/socket interface and vertical stiffness as indicators, changes in material properties, socket alignment and socket rectification are investigated; assumptions about the frictional characteristic at the interface are seen to be critical in determination of load distribution. This type of analysis may provide the next stage of refinement for computer-aided socket design systems.

Entities:  

Mesh:

Year:  1992        PMID: 1487943     DOI: 10.1007/bf02446180

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  6 in total

1.  Pressures in critical regions of the below-knee patellar-tendon-bearing prosthesis.

Authors:  J R Pearson; G Holmgren; L March; K Oberg
Journal:  Bull Prosthet Res       Date:  1973

2.  Stump-socket fit of below-knee prostheses: comparison of three methods of measurement.

Authors:  R H Meier; E D Meeks; R M Herman
Journal:  Arch Phys Med Rehabil       Date:  1973-12       Impact factor: 3.966

3.  Effect of liner materials on interface pressures in below-knee prostheses.

Authors:  W A Sonck; J L Cockrell; G H Koepke
Journal:  Arch Phys Med Rehabil       Date:  1970-11       Impact factor: 3.966

Review 4.  Artificial sensing skin mimicking mechanoelectrical conversion properties of human dermis.

Authors:  D De Rossi; A Nannini; C Domenici
Journal:  IEEE Trans Biomed Eng       Date:  1988-02       Impact factor: 4.538

5.  Computer aided design of prosthetic sockets for below-knee amputees.

Authors:  C G Saunders; J Foort; M Bannon; D Lean; L Panych
Journal:  Prosthet Orthot Int       Date:  1985-04       Impact factor: 1.895

6.  An analysis of soft tissue loading in the foot--a preliminary report.

Authors:  S Nakamura; R D Crowninshield; R R Cooper
Journal:  Bull Prosthet Res       Date:  1981
  6 in total
  3 in total

Review 1.  Interface mechanics in external prosthetics: review of interface stress measurement techniques.

Authors:  J E Sanders
Journal:  Med Biol Eng Comput       Date:  1995-07       Impact factor: 2.602

2.  Effect of Local Neck Anatomy on Localized One-Dimensional Measurements of Arterial Stiffness: A Finite-Element Model Study.

Authors:  Adriaan Campo; Matthew D McGarry; Thomas Panis; Joris Dirckx; Elisa Konofagou
Journal:  J Biomech Eng       Date:  2019-03-01       Impact factor: 2.097

3.  Analysis of Pressure Distribution in Transfemoral Prosthetic Socket for Prefabrication Evaluation via the Finite Element Method.

Authors:  Mohd Syahmi Jamaludin; Akihiko Hanafusa; Yamamoto Shinichirou; Yukio Agarie; Hiroshi Otsuka; Kengo Ohnishi
Journal:  Bioengineering (Basel)       Date:  2019-10-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.