| Literature DB >> 1476416 |
L C Danford1, D A Schoeller, R F Kushner.
Abstract
We compared two bioelectrical impedance analysis models, the right-sided tetrapolar method and an experimental cylindrical method, to total body water (TBW) determined by isotopic dilution in 37 prepubertal children aged 5-9 years, who were divided into development and cross-validation groups. Despite its theoretical advantage, no improvement in predicted TBW was seen with the experimental method. The best predictive equation, using the tetrapolar method, was TBW (kg) = 1.84 + 0.45 (height square/resistance) +0.11(weight); r2 = 0.98; SEE = 0.62 kg.Entities:
Mesh:
Year: 1992 PMID: 1476416 DOI: 10.1080/03014469200002422
Source DB: PubMed Journal: Ann Hum Biol ISSN: 0301-4460 Impact factor: 1.533