| Literature DB >> 14763744 |
J Flarakos1, W Luo, M Aman, D Svinarov, N Gerber, P Vouros.
Abstract
A robust and validated LC-MS-MS quantitative method, using column switching and mutiple reaction monitoring was developed for the analysis of risperidone (RIS) and 9-hydroxyrisperidone in human plasma and saliva. The analytical range was 1-100 ng/ml. The method used 25 microl of sample precipitated with 75 microl of acetonitrile containing internal standard (R068808). Analyses were conducted on a PE Sciex API-III + triple quadrupole mass spectrometer fitted with a Turbo IonSpray source. The method was validated for human plasma using EDTA as the anticoagulant and cross-validated to heparinized human plasma and saliva. The recoveries of risperidone and 9-hydroxyrisperidone were 90-93 and 89-93%, respectively. The validated method was applied to clinical samples to study risperidone and 9-hydroxyrisperidone concentrations in plasma and saliva. Risperidone and 9-hydroxyrisperidone appear in the saliva of patients treated with risperidone. Their detection/quantification in saliva provides evidence for recent adherence with therapy.Entities:
Mesh:
Substances:
Year: 2004 PMID: 14763744 DOI: 10.1016/j.chroma.2003.10.138
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759