Endovascular rescue of a maldeployed aortic stent-graft causing renal artery occlusion: technical considerations.

Peter H Lin1, Ruth L Bush, Alan B Lumsden.   

Abstract

The technical success of an endovascular abdominal aortic aneurysm repair depends on the endograft exclusion of the aneurysm along with preservation of flow to major aortic branches, such as the renal or hypogastric artery. A proximally malpositioned aortic endograft may result in inadvertent renal artery occlusion, which may lead to renal infarction with irreversible loss of the renal function. In this report, the authors describe an endovascular rescue technique to salvage a proximally maldeployed aortic endograft that inadvertently covered the renal artery orifice. Physicians performing endovascular aneurysm procedures should be aware of such a salvage maneuver in the event an aortic endograft is proximally maldeployed and occludes the renal artery orifice.

Entities:  

Mesh:

Year:  2004        PMID: 14760480     DOI: 10.1177/153857440403800109

Source DB:  PubMed          Journal:  Vasc Endovascular Surg        ISSN: 1538-5744            Impact factor:   1.089