Effects of stent stiffness on local haemodynamics with particular reference to wave reflections.

H Alderson1, M Zamir.   

Abstract

The placement of a rigid stent within an elastic vessel produces wave reflection sites at the entrance to and exit from the stent. The net haemodynamic effects of these reflections depend critically on the degree of stiffness of the stent and on its length and position within the diseased vessel, variables that have been found to affect the clinical performance of a stent. Here these effects are examined analytically, using a segmented tube model. The results indicate that the presence of the stent within the larger diseased vessel has the effect of producing higher pressure at the vessel entrance than that at exit. This pressure difference, when superimposed on the underlying pressure distribution within the vessel, has the net effect of actually aiding rather than impeding the flow, but the extent of this depends on the length and position of the stent. A short stent placed near the entrance of the diseased vessel may be favoured clinically for producing the least perturbation in the underlying haemodynamics and thus reducing the chance of restenosis, while a long stent placed near the exit may be favoured for producing a positive pressure difference and thus aiding the flow.

Entities:  

Mesh:

Year:  2004        PMID: 14757453     DOI: 10.1016/s0021-9290(03)00289-6

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  8 in total

Review 1.  Clinical significance and technical assessment of stent cell geometry in carotid artery stenting.

Authors:  Gail M Siewiorek; Ender A Finol; Mark H Wholey
Journal:  J Endovasc Ther       Date:  2009-04       Impact factor: 3.487

2.  Cerebral hemodynamics and intracranial aneurysms: Reflecting on pipeline embolization devices.

Authors:  Wesley K Lefferts; Kevin S Heffernan
Journal:  Interv Neuroradiol       Date:  2018-07-22       Impact factor: 1.610

3.  Pulmonary artery capacitance and pulmonary vascular resistance as prognostic indicators in acute pulmonary embolism.

Authors:  Juan A Quintero-Martinez; Waldemar E Wysokinski; Sandra N Cordova-Madera; Renzo J Mogollon; Mariana Garcia-Arango; Danielle T Vlazny; Damon E Houghton; Ana I Casanegra; Hector R Villarraga
Journal:  Eur Heart J Open       Date:  2022-02-23

Review 4.  Medical Textiles as Vascular Implants and Their Success to Mimic Natural Arteries.

Authors:  Charanpreet Singh; Cynthia S Wong; Xungai Wang
Journal:  J Funct Biomater       Date:  2015-06-30

5.  Compliance Study of Endovascular Stent Grafts Incorporated with Polyester and Polyurethane Graft Materials in both Stented and Unstented Zones.

Authors:  Ying Guan; Lu Wang; Jing Lin; Martin W King
Journal:  Materials (Basel)       Date:  2016-08-05       Impact factor: 3.623

6.  Wave reflection and transmission in multiply stented blood vessels.

Authors:  T K Papathanasiou; A B Movchan; D Bigoni
Journal:  Proc Math Phys Eng Sci       Date:  2017-06-07       Impact factor: 2.704

7.  Left Main Stenting Induced Flow Disturbances on Ascending Aorta and Aortic Arch.

Authors:  Gianluca Rigatelli; Marco Zuin; Alan Fong; Truyen Ttt Tai; Thach Nguyen
Journal:  J Transl Int Med       Date:  2019-03-29

8.  Reflected wave intensity increases based on aortic diameter after endovascular aortic therapy in a goat model.

Authors:  Tomohiro Takano; Masumi Iwai-Takano; Yusuke Tsuboko; Yasuyuki Shiraishi; Tomoyuki Yambe; Takashi Igarashi; Hitoshi Yokoyama
Journal:  Sci Rep       Date:  2021-02-15       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.