Ultrasound imaging of milk ejection in the breast of lactating women.

Donna T Ramsay1, Jacqueline C Kent, Robyn A Owens, Peter E Hartmann.   

Abstract

OBJECTIVE: Currently, the methods for assessing milk ejection in women include serial sampling of plasma oxytocin and measurement of intraductal pressure, both of which are invasive and may induce stress. We hypothesized that milk ejection would cause an increase in milk-duct diameter that could be observed noninvasively with ultrasound, and this could be used to investigate the physiology of milk ejection in women.
METHODS: One milk duct was scanned in the unsuckled breast in 2 groups of mothers: group BB (n = 21) for the beginning of a breastfeed and group EB (n = 24) for the entire breastfeed. A duct also was monitored for a 5-minute period on 2 separate days in the absence of factors that may induce milk ejection in group EB to provide a baseline duct diameter. Milk intake at a breastfeed was measured by test weighing.
RESULTS: A significant increase in milk-duct diameter was observed when milk ejection was sensed and/or the infant changed its swallowing pattern in both groups. Multiple increases and decreases (mean: 2.5 per breastfeed; standard deviation: 1.5; n = 62) in duct diameter occurred in group EB. Duct diameter remained relatively stable between breastfeeds (coefficient of variation: 1.4%-8.3%). Infant milk intake was positively related to the number of milk ejections (r2 =.365; n = 57).
CONCLUSIONS: Ultrasound is an objective, noninvasive technique for detecting milk ejection by observing an increase in milk-duct diameter. However, this technique requires an experienced ultrasonographer, adequate imaging time, and surroundings conducive to breastfeeding. Multiple milk ejections were common during breastfeeding, although they were not sensed by mothers. The number of milk ejections influenced the amount of milk the infant consumed.

Entities:  

Mesh:

Year:  2004        PMID: 14754950     DOI: 10.1542/peds.113.2.361

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  74 in total

Review 1.  The origin of human milk bacteria: is there a bacterial entero-mammary pathway during late pregnancy and lactation?

Authors:  Juan M Rodríguez
Journal:  Adv Nutr       Date:  2014-11-14       Impact factor: 8.701

2.  Anatomy of the lactating human breast redefined with ultrasound imaging.

Authors:  D T Ramsay; J C Kent; R A Hartmann; P E Hartmann
Journal:  J Anat       Date:  2005-06       Impact factor: 2.610

Review 3.  Immune cell-mediated protection of the mammary gland and the infant during breastfeeding.

Authors:  Foteini Hassiotou; Donna T Geddes
Journal:  Adv Nutr       Date:  2015-05-15       Impact factor: 8.701

Review 4.  Differential Establishment of Bifidobacteria in the Breastfed Infant Gut.

Authors:  Zachery T Lewis; David A Mills
Journal:  Nestle Nutr Inst Workshop Ser       Date:  2017-03-27

5.  Microbiota of human breast tissue.

Authors:  Camilla Urbaniak; Joanne Cummins; Muriel Brackstone; Jean M Macklaim; Gregory B Gloor; Chwanrow K Baban; Leslie Scott; Deidre M O'Hanlon; Jeremy P Burton; Kevin P Francis; Mark Tangney; Gregor Reid
Journal:  Appl Environ Microbiol       Date:  2014-03-07       Impact factor: 4.792

6.  Simultaneous breast expression in breastfeeding women is more efficacious than sequential breast expression.

Authors:  Danielle K Prime; Catherine P Garbin; Peter E Hartmann; Jacqueline C Kent
Journal:  Breastfeed Med       Date:  2012-10-05       Impact factor: 1.817

7.  Role of breast magnetic resonance imaging (MRI) in patients with unilateral nipple discharge: preliminary study.

Authors:  L Ballesio; C Maggi; S Savelli; M Angeletti; C De Felice; M L Meggiorini; L Manganaro; L M Porfiri
Journal:  Radiol Med       Date:  2008-04-02       Impact factor: 3.469

8.  Ultrasound imaging of infant swallowing during breast-feeding.

Authors:  Donna T Geddes; Lynda M Chadwick; Jacqueline C Kent; Catherine P Garbin; Peter E Hartmann
Journal:  Dysphagia       Date:  2009-07-22       Impact factor: 3.438

9.  Using milk flow rate to investigate milk ejection in the left and right breasts during simultaneous breast expression in women.

Authors:  Danielle K Prime; Donna T Geddes; Diane L Spatz; Marc Robert; Naomi J Trengove; Peter E Hartmann
Journal:  Int Breastfeed J       Date:  2009-10-26       Impact factor: 3.461

10.  The use of ultrasound to identify milk ejection in women - tips and pitfalls.

Authors:  Donna T Geddes
Journal:  Int Breastfeed J       Date:  2009-06-01       Impact factor: 3.461

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.