Abstract

OBJECTIVE: We sought to evaluate the safety and efficacy of the first available totally implantable replacement heart (AbioCor Implantable Replacement Heart System) in the treatment of severe, irreversible biventricular heart failure in human patients.
METHODS: Seven male adult patients with severe, irreversible biventricular failure (>70% thirty-day predicted mortality) who were not candidates for transplantation met all institutional review board study criteria and had placement of the AbioCor Implantable Replacement Heart. All were in cardiogenic shock despite maximal medical therapy, including inotropes and intra-aortic balloon pumps. Mean age was 66.7 +/- 10.4 years (range, 51-79 years). Four of 7 patients had prior operations. Six had ischemic and one had idiopathic cardiomyopathy. All had 3-dimensional computer-simulated implantation of the thoracic unit that predicted adequate fit. At the time of the operation, the internal transcutaneous energy transfer coil, battery, and controller were placed. Biventriculectomy was then performed, and the thoracic unit was placed in an orthotopic position and attached to the atrial cuffs and outflow conduits with quick-connects. The flow was adjusted to 4 to 8 L/min. Central venous and left atrial pressures were maintained at 5 to 15 mm Hg. The device is powered through transcutaneous energy transfer. An atrial flow-balancing chamber is used to adjust left/right balance. The balance chamber and transcutaneous energy transfer eliminate the need for percutaneous lines.
RESULTS: There was one intraoperative death caused by coagulopathic bleeding and one early death caused by an aprotinin reaction. There have been multiple morbidities primarily related to preexisting illness severity: 5 patients had prolonged intubation, 2 had hepatic failure (resolved in 1), 4 had renal failure (resolved in 3), and 1 each had recurrent gastrointestinal bleeding, acute cholecystitis requiring laparotomy, respiratory failure that resolved after 3 days of extracorporeal membrane oxygenation, and malignant hyperthermia (resolved). There were 3 late deaths: one caused by multiple systems organ failure (postoperative day 56), one caused by a cerebrovascular accident (postoperative day 142), and one caused by retroperitoneal bleeding and resultant multiple systems organ failure (postoperative day 151). This latter patient was not able to tolerate anticoagulation (no anticoagulation or antiplatelet therapy alone for 80% of the first 60 days) and had a transient ischemic attack on postoperative day 61 and a cerebrovascular accident on postoperative day 130. At autopsy, blood pumps were clean. The 2 patients who had large cerebrovascular accidents had thrombus on the atrial cage struts. These struts have been removed for future implants. There has been no significant hemolysis or device-related infections. The balance chamber has allowed for left/right balance in all patients (left atrial pressure within 5 mm Hg of right atrial pressure). Three patients have taken multiple (>50) trips out of the hospital, and 2 have been discharged from the hospital. Total days on support with the AbioCor are 759.
CONCLUSION: The initial clinical experience suggests that the AbioCor might be effective therapy in patients with advanced biventricular failure. There have been no significant device malfunctions. Two of these patients have been discharged from the hospital.

Entities:  

Mesh:

Year:  2004        PMID: 14752423     DOI: 10.1016/j.jtcvs.2003.07.023

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  27 in total

Review 1.  Mechanical circulatory support devices (MCSD) in Japan: current status and future directions.

Authors:  Setsuo Takatani; Hikaru Matsuda; Akihisa Hanatani; Chisato Nojiri; Kenji Yamazaki; Tadashi Motomura; Katsuhiro Ohuchi; Tohru Sakamoto; Takashi Yamane
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

Review 2.  Surgical aspects of congestive heart failure.

Authors:  Daniel J Goldstein; Douglas Smego; Robert E Michler
Journal:  Heart Fail Rev       Date:  2006-06       Impact factor: 4.214

Review 3.  Mechanical circulatory support.

Authors:  Deborah J Kozik; Mark D Plunkett
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

4.  Hydrodynamic characteristics of the helical flow pump.

Authors:  Kohei Ishii; Kyohei Hosoda; Masahiro Nishida; Takashi Isoyama; Itsuro Saito; Koki Ariyoshi; Yusuke Inoue; Toshiya Ono; Hidemoto Nakagawa; Masami Sato; Sintaro Hara; Xinyang Lee; Sheng-Yuan Wu; Kou Imachi; Yusuke Abe
Journal:  J Artif Organs       Date:  2015-03-18       Impact factor: 1.731

Review 5.  Total artificial hearts: past, present, and future.

Authors:  William E Cohn; Daniel L Timms; O H Frazier
Journal:  Nat Rev Cardiol       Date:  2015-06-02       Impact factor: 32.419

6.  Antibacterial Biomimetic Hybrid Films.

Authors:  M Carme Coll Ferrer; Noreen J Hickok; David M Eckmann; Russell J Composto
Journal:  Soft Matter       Date:  2013-02-28       Impact factor: 3.679

Review 7.  An insight into short- and long-term mechanical circulatory support systems.

Authors:  Markus Ferrari; Peter Kruzliak; Kyriakos Spiliopoulos
Journal:  Clin Res Cardiol       Date:  2014-10-28       Impact factor: 5.460

8.  Managing drugs and devices in patients with permanent ventricular assist devices.

Authors:  Martin Cadeiras; Manuel von Bayern; Mario C Deng
Journal:  Curr Treat Options Cardiovasc Med       Date:  2007-08

9.  Progress on the design and development of the continuous-flow total artificial heart.

Authors:  Mariko Kobayashi; David J Horvath; Nicole Mielke; Akira Shiose; Barry Kuban; Mark Goodin; Kiyotaka Fukamachi; Leonard A R Golding
Journal:  Artif Organs       Date:  2012-07-02       Impact factor: 3.094

10.  An innovative, sensorless, pulsatile, continuous-flow total artificial heart: device design and initial in vitro study.

Authors:  Kiyotaka Fukamachi; David J Horvath; Alex L Massiello; Hideyuki Fumoto; Tetsuya Horai; Santosh Rao; Leonard A R Golding
Journal:  J Heart Lung Transplant       Date:  2009-09-26       Impact factor: 10.247

View more