Literature DB >> 14752309

A new, wider-diameter Crawford tube for stenting in the lacrimal drainage system.

Jeffrey J Hurwitz1.   

Abstract

PURPOSE: To determine whether technology is available to secure a larger-diameter Crawford tube swedged onto the metal Crawford introducer and to use the system in a closed intubation to treat a failed dacryocystorhinostomy.
METHODS: A metal ferrule has been developed to swedge a larger-diameter Crawford tube onto the standard Crawford probe. Seven patients with failed dacryocystorhinostomy procedures were treated with probing and intubation with the use of this larger-tube Crawford system.
RESULTS: Six of the 7 patients had patent lacrimal systems after the tubes were removed, a result comparable to open revision after a failed dacryocystorhinostomy. In each case, on insertion, the tube remained firmly attached to the Crawford probe and did not dislodge.
CONCLUSIONS: The technology of using a metal ferrule at the junction of the standard Crawford probe and a larger silicone tube was successful in creating a Crawford system with greater dilation capability than the standard Crawford system. The technique of probing and intubation with this larger Crawford system may be a simple alternative to open revision of a failed dacryocystorhinostomy procedure.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14752309     DOI: 10.1097/01.IOP.0000103008.01277.BE

Source DB:  PubMed          Journal:  Ophthalmic Plast Reconstr Surg        ISSN: 0740-9303            Impact factor:   1.746


  3 in total

1.  Bicanalicular double silicone stenting in endoscopic dacryocystorhinostomy with lacrimal trephination in distal or common canalicular obstruction.

Authors:  Ji-Sun Paik; Won-Kyung Cho; Suk-Woo Yang
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-12-01       Impact factor: 2.503

2.  Comparison of the Efficacies of 0.94 mm and Double Silicone Tubes for Treatment of Canalicular Obstruction.

Authors:  Seong Chan Choi; Hye Sun Choi; Jae Woo Jang; Sung Joo Kim; Jung Hye Lee
Journal:  Korean J Ophthalmol       Date:  2017-02-02

3.  Comparison of the efficacies of 1.0 and 1.5 mm silicone tubes for the treatment of nasolacrimal duct obstruction.

Authors:  Jutaro Nakamura; Tomoyuki Kamao; Arisa Mitani; Nobuhisa Mizuki; Atsushi Shiraishi
Journal:  Sci Rep       Date:  2022-07-11       Impact factor: 4.996

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.