Experimental study of quantitative assessment of left ventricular mass with contrast enhanced real-time three-dimensional echocardiography.

Lei Zhuang1, Xin-Fang Wang, Ming-Xing Xie, Li-Xin Chen, Hong-Wen Fei, Ying Yang, Jing Wang, Run-Qing Huang, Ou-Di Chen, Liang-Yu Wang.   

Abstract

OBJECTIVES: To evaluate the feasibility and accuracy of measurement of left ventricular mass with intravenous contrast enhanced real-time three-dimensional (RT3D) echocardiography in the experimental setting.
METHODS: RT3D echocardiography was performed in 13 open-chest mongrel dogs before and after intravenous infusion of a perfluorocarbon contrast agent. Left ventricular myocardium volume was measured according to the apical four-plane method provided by TomTec 4D cardio-View RT1.0 software, then the left ventricular mass was calculated as the myocardial volume multiplied by the relative density of myocardium. Correlative analysis and paired t-test were performed between left ventricular mass obtained from RT3D echocardiography and the anatomic measurements.
RESULTS: Anatomic measurement of total left ventricular mass was 55.6 +/- 9.3 g, whereas RT3D echocardiographic calculation of left ventricular mass before and after intravenous perfluorocarbon contrast agent was 57.5 +/- 11.4 and 55.5 +/- 9.3 g, respectively. A significant correlation was observed between the RT3D echocardiographic estimates of total left ventricular mass and the corresponding anatomic measurements (r = 0.95). A strong correlation was found between RT3D echocardiographic estimates of left ventricular mass with perfluorocarbon contrast and the anatomic results (r = 0.99). Analysis of intraobserver and interobserver variability showed strong indexes of agreement in the measurement of left ventricular mass with pre and post-contrast RT3D echocardiography.
CONCLUSIONS: Measurements of left ventricular mass derived from RT3D echocardiography with and without intravenous contrast showed a significant correlation with the anatomic results. Contrast enhanced RT3D echocardiography permitted better visualization of the endocardial border, which would provide a more accurate and reliable means of determining left ventricular myocardial mass in the experimental setting.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14750411

Source DB:  PubMed          Journal:  J Cardiol        ISSN: 0914-5087            Impact factor:   3.159


  2 in total

1.  Vascular lesions and s-thrombomodulin concentrations from auricular arteries of rabbits infused with microbubble contrast agent and exposed to pulsed ultrasound.

Authors:  James F Zachary; James P Blue; Rita J Miller; William D O'Brien
Journal:  Ultrasound Med Biol       Date:  2006-11       Impact factor: 2.998

Review 2.  Real-Time Three-Dimensional Echocardiography: Characterization of Cardiac Anatomy and Function-Current Clinical Applications and Literature Review Update.

Authors:  Omar Velasco; Morgan Q Beckett; Aaron W James; Megan N Loehr; Taylor G Lewis; Tahmin Hassan; Rajesh Janardhanan
Journal:  Biores Open Access       Date:  2017-03-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.