| Literature DB >> 14744000 |
Rainer A Leitgeb1, Leopold Schmetterer, Christoph K Hitzenberger, Adolf F Fercher, Fatma Berisha, Maciej Wojtkowski, Tomasz Bajraszewski.
Abstract
The possibility of measuring a full Doppler flow depth profile in parallel by use of frequency-domain optical coherence tomography is demonstrated. The method is based on a local phase analysis of the backscattered signal and allows for imaging of bidirectional Doppler flow. The Doppler frequency limit is 5 kHz for the presented measurements and is set by half of the frame rate of the CCD detector array. We measured the flow of 0.3-microm microspheres suspended in distilled water at controlled flow rates and in vitro human blood flow through a 200-microm capillary with a real-time color-encoded Doppler tomogram rate of 2-3/s.Entities:
Mesh:
Year: 2004 PMID: 14744000 DOI: 10.1364/ol.29.000171