Literature DB >> 14743926

Performance of bipolar forceps during coagulation and its dependence on the tip material: a quantitative experimental assay. Technical note.

Takeshi Mikami1, Akira Takahashi, Kazuo Hashi, Shinsei Gasa, Kiyohiro Houkin.   

Abstract

The aim of this study was to measure objectively the adherence of burned tissue to bipolar forceps to evaluate the coagulation performance of forceps made of different types of metals. Coagulation performance of bipolar forceps made of gold, titanium, and stainless steel was determined by comparing the amount of protein in the adhered coagulum on the tips. The amount of adhered coagulum was significantly less on the gold-plated bipolar forceps than on those made of the other two materials. The ease with which coagulum could be removed was compared using the cleaning cycle of an ultrasonic rinsing device. This ease of removal was also significant with the gold-plated forceps. Electron microscopy observations of the surface of the forceps tips revealed a significant difference in roughness among these materials, and there were also significant differences in wetting tensions. Measuring adherence based on three different types of roughness and wetting tensions of forceps made from the same metal (titanium) also demonstrated a significant difference in the cleaning cycle. Histological examination of an artery coagulated with the gold-plated bipolar forceps showed that the structure had been completely collapsed without destruction of the layers, whereas arteries coagulated with the other materials revealed severely damaged structures. Adherence to bipolar forceps was dependent on both the material in the tips and the roughness of this material. The gold-plated bipolar forceps demonstrated the best performance.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14743926     DOI: 10.3171/jns.2004.100.1.0133

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  9 in total

1.  Bipolar vessel sealing: instrument contamination and wear have little effect on seal quality and success in a porcine in vitro model.

Authors:  Christian W Wallwiener; Susanna H Junginger; Wolfgang Zubke; Sara Y Brucker; Markus D Enderle; Alexander Neugebauer; Birgitt Schönfisch; Markus Wallwiener
Journal:  Langenbecks Arch Surg       Date:  2014-07-27       Impact factor: 3.445

2.  Biomedical electrosurgery devices containing nanostructure for minimally invasive surgery: reduction of thermal injury and acceleration of wound healing for liver cancer.

Authors:  Wen-Tien Hsiao; Li-Hsiang Lin; Hsi-Jen Chiang; Keng-Liang Ou; Han-Yi Cheng
Journal:  J Mater Sci Mater Med       Date:  2015-01-29       Impact factor: 3.896

3.  Ultrasound-based relative elastic modulus imaging for visualizing thermal ablation zones in a porcine model.

Authors:  Jingfeng Jiang; Chris Brace; Anita Andreano; Ryan J DeWall; Nick Rubert; Ted G Fisher; Tomy Varghese; Fred Lee; Timothy J Hall
Journal:  Phys Med Biol       Date:  2010-03-30       Impact factor: 3.609

4.  Biomedical nanostructured coating for minimally invasive surgery devices applications: characterization, cell cytotoxicity evaluation and an animal study in rat.

Authors:  Keng-Liang Ou; Jan-Show Chu; Hossein Hosseinkhani; Jeng-Fong Chiou; Chih-Hua Yu
Journal:  Surg Endosc       Date:  2014-03-12       Impact factor: 4.584

5.  Effect of nanostructured thin film on minimally invasive surgery devices applications: characterization, cell cytotoxicity evaluation and an animal study in rat.

Authors:  Keng-Liang Ou; Chao-Chia Weng; Erwan Sugiatno; Muhammad Ruslin; Yun-Ho Lin; Han-Yi Cheng
Journal:  Surg Endosc       Date:  2015-11-12       Impact factor: 4.584

6.  Young's modulus reconstruction for radio-frequency ablation electrode-induced displacement fields: a feasibility study.

Authors:  Jingfeng Jiang; Tomy Varghese; Christopher L Brace; Ernest L Madsen; Timothy J Hall; Shyam Bharat; Maritza A Hobson; James A Zagzebski; Fred T Lee
Journal:  IEEE Trans Med Imaging       Date:  2009-02-27       Impact factor: 10.048

7.  Comparatıve assessment of thermal ınjury ınduced by bıpolar electrocautery systems ın a porcıne model.

Authors:  Anil K Roy; Nefize Turan; Pasang Wangmo; Louis Nkrumah; Stewart G Neill; Gustavo Pradilla
Journal:  Surg Neurol Int       Date:  2021-04-08

8.  Evaluation of Bone Wax Coated Bipolar Coagulation Forceps: Performance and Safety Assessment.

Authors:  Jichun Shi; Wei Wei; Zhen Wang; Haobin Ren; Chenguang Jia; Lixin Dong; Zhengwei Li; Jianjian Zhang; Yu Feng; Kaixin Huang; Xiang Li; Jincao Chen
Journal:  Front Surg       Date:  2022-01-20

9.  Comparison of thermal coagulation profiles for bipolar forceps with different cooling mechanisms in a porcine model of spinal surgery.

Authors:  Roland K Chen; Khoi D Than; Anthony C Wang; Paul Park; Albert J Shih
Journal:  Surg Neurol Int       Date:  2013-09-06
  9 in total