Literature DB >> 14742964

Long-term outcome of spinal cord stimulation and hardware complications.

D Gavin Quigley1, Jonathan Arnold, Paul R Eldridge, Heather Cameron, Kate McIvor, John B Miles, T R K Varma.   

Abstract

Spinal cord stimulation (SCS) is a treatment modality for medically intractable chronic pain. This study reports an 11-year experience with SCS assessing long-term pain relief and specifically evaluating complications and revisions. It took the form of a retrospective review of medical/surgical records with a postal questionnaire. The subjects were 102 patients with medically intractable chronic pain who underwent SCS implantation between 1989 and 2000. There were 64 revision operations carried out on 35 patients. These comprised electrode replacement/repositioning (29), generator replacement (23), cable failure (3) and implant removal (5). Five (4.9%) implants became infected and 2 required removal. Clinician-reported pain relief was substantial in 69 (68%) patients. This study adds to the weight of evidence that patients undergoing SCS derive significant benefits in terms of pain relief. However, revision rates remain high due to technical and biological factors. Copyright 2003 S. Karger AG, Basel

Entities:  

Mesh:

Year:  2003        PMID: 14742964     DOI: 10.1159/000075104

Source DB:  PubMed          Journal:  Stereotact Funct Neurosurg        ISSN: 1011-6125            Impact factor:   1.875


  6 in total

1.  Midline minimally invasive placement of spinal cord stimulators: a technical note.

Authors:  Edison P Valle-Giler; Wale A R Sulaiman
Journal:  Ochsner J       Date:  2014

Review 2.  Spinal cord stimulation: an update.

Authors:  Steven Falowski; Amanda Celii; Ashwini Sharan
Journal:  Neurotherapeutics       Date:  2008-01       Impact factor: 7.620

3.  Nanoscale neuroelectrode modification via sub-20 nm silicon nanowires through self-assembly of block copolymers.

Authors:  Parvaneh Mokarian-Tabari; Catalina Vallejo-Giraldo; Marc Fernandez-Yague; Cian Cummins; Michael A Morris; Manus J P Biggs
Journal:  J Mater Sci Mater Med       Date:  2015-02-13       Impact factor: 3.896

4.  Spontaneous lead breakage in implanted spinal cord stimulation systems.

Authors:  Tae Hun Kim; Pyung Bok Lee; Hye Min Son; Jong Bum Choi; Jee Youn Moon
Journal:  Korean J Pain       Date:  2010-03-10

5.  Dorsal column stimulator applications.

Authors:  Claudio Yampolsky; Santiago Hem; Damián Bendersky
Journal:  Surg Neurol Int       Date:  2012-10-31

6.  Brain-spine interfaces to reverse paralysis.

Authors:  Henri Lorach; Guillaume Charvet; Jocelyne Bloch; Grégoire Courtine
Journal:  Natl Sci Rev       Date:  2022-01-18       Impact factor: 23.178

  6 in total